Assessment of endoscope reprocessing at World Gastroenterology Organisation training centers using adenosine triphosphate testing.
Autor: | Visrodia K; Division of Digestive and Liver Diseases, Columbia University Irving Medical Center - New York Presbyterian Hospital, New York, USA., Jones D; Department of Gastroenterology & Hepatology, Logan Hospital, Queensland, Australia., Jennings M; Department of Gastroenterology and Hepatology, Royal Perth Hospital, Perth, Western Australia, Australia., Conway C; Gastroenterological Nurses College of Australia, New South Wales, Australia., Burgos H; Gastroclinica WGO Training Center, San Jose, Costa Rica., Malani J; School of Public Health and Primary Care, Fiji National University, Suva, Fiji., Ogutu E; Department of Internal Medicine, University of Nairobi, Nairobi, Kenya, Africa., Pausawasdi N; Siriraj GI Endoscopy Center and Division of Gastroenterology, Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand., Sabbagh L; Department of Gastroenterology, Reina Sofia Clinic and Colombia University Clinic, Bogota, Colombia., Leddin D; Department of Medicine, Dalhousie University, Halifax, Canada., Kassmeyer B; Division of Biomedical Statistics and Informatics, Mayo Clinic, Rochester, United States., Lennon RJ; Division of Biomedical Statistics and Informatics, Mayo Clinic, Rochester, United States., Topazian MD; Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, United States. |
---|---|
Jazyk: | angličtina |
Zdroj: | Endoscopy international open [Endosc Int Open] 2024 Jan 08; Vol. 12 (1), pp. E68-E77. Date of Electronic Publication: 2024 Jan 08 (Print Publication: 2024). |
DOI: | 10.1055/a-2185-6035 |
Abstrakt: | Background and study aims Adequacy of endoscope disinfection in resource-limited settings is unknown. Adenosine triphosphate (ATP) testing is useful for evaluation of endoscope reprocessing, and ATP <200 relative light units (RLUs) after manual endoscope cleaning has been associated with adequacy of endoscope disinfection. Methods Consecutive endoscopes undergoing reprocessing at five World Gastroenterology Organisation (WGO) training centers underwent ATP testing before and after an on-site educational intervention designed to optimize reprocessing practices. Results A total of 343 reprocessing cycles of 65 endoscopes were studied. Mean endoscope age was 5.3 years (range 1-13 years). Educational interventions, based on direct observation of endoscope reprocessing practices at each site, included refinements in pre-cleaning, manual cleaning, high-level disinfection, and endoscope drying and storage. The percentage of reprocessing cycles with post-manual cleaning ATP ≧200 decreased from 21.4% prior to educational intervention to 14.8% post-intervention ( P =0.11). In multivariable logistic modelling, gastroscopes were significantly less likely (odds ratio [OR] 0.04, 95% confidence interval [CI] 0.01-0.19; P <0.001) than colonoscopes to achieve post-manual cleaning ATP < 200. No other factor (educational intervention, study site, endoscope age) was significantly associated with improved outcomes. Endoscope ID was not significantly associated with ATP values, and sites that performed manual versus automated HLD did not have significantly different likelihood of post-manual cleaning ATP <200 (OR 1.18, 95% CI 0.56-2.50; P =0.67). Conclusions In resource-limited settings, approximately 20% of endoscope reprocessing cycles may result in inadequate disinfection. This was not significantly improved by a comprehensive educational intervention. Alternative approaches to endoscope reprocessing are needed. Competing Interests: Conflict of Interest The authors declare that they have no conflict of interest. (The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).) |
Databáze: | MEDLINE |
Externí odkaz: |