Prospective Comparison of Handheld Ultrasound Devices from Different Manufacturers with Respect to B-Scan Quality and Clinical Significance for Various Abdominal Sonography Questions.

Autor: Merkel D; BIKUS-Brandenburg Institute for Clinical Ultrasound, Brandenburg Medical School Theodor Fontane (MHB), 16816 Neuruppin, Germany.; Immanuel Klinik Rüdersdorf, University Hospital of the Brandenburg Medical School, 15562 Rüdersdorf bei Berlin, Germany., Züllich TF; Immanuel Klinik Rüdersdorf, University Hospital of the Brandenburg Medical School, 15562 Rüdersdorf bei Berlin, Germany., Schneider C; Immanuel Klinik Rüdersdorf, University Hospital of the Brandenburg Medical School, 15562 Rüdersdorf bei Berlin, Germany., Yousefzada M; Immanuel Klinik Rüdersdorf, University Hospital of the Brandenburg Medical School, 15562 Rüdersdorf bei Berlin, Germany., Beer D; Immanuel Klinik Rüdersdorf, University Hospital of the Brandenburg Medical School, 15562 Rüdersdorf bei Berlin, Germany., Ludwig M; Department of Internal Medicine I, Hospital of the German Armed Forces Berlin, 10115 Berlin, Germany., Weimer A; Center of Orthopedics, Trauma Surgery and Spinal Cord Injury, Heidelberg University Hospital Heidelberg, 69118 Heidelberg, Germany., Künzel J; Department of Otorhinolaryngology, Head and Neck Surgery, University Hospital Regensburg, 93053 Regensburg, Germany., Kloeckner R; Institute of Interventional Radiology, University Hospital Schleswig-Holstein-Campus Lübeck, 23538 Lübeck, Germany., Weimer JM; Rudolf Frey Learning Clinic, University Medical Center of the Johannes Gutenberg University Mainz, 55131 Mainz, Germany.
Jazyk: angličtina
Zdroj: Diagnostics (Basel, Switzerland) [Diagnostics (Basel)] 2023 Dec 08; Vol. 13 (24). Date of Electronic Publication: 2023 Dec 08.
DOI: 10.3390/diagnostics13243622
Abstrakt: Background: Handheld ultrasound (HHUS) devices have chiefly been deployed in emergency medicine, where they are considered a valid tool. The data situation is less clear in the case of internal questions in abdominal sonography. In our study, we investigate whether HHUS devices from different manufacturers differ in their B-scan quality, and whether any differences are relevant for the significance of an internal ultrasound examination.
Method: The study incorporated eight HHUS devices from different manufacturers. Ultrasound videos of seven defined sonographic questions were recorded with all of the devices. The analogue recording of the same findings with a conventional high-end ultrasound (HEUS) device served as an evaluation criterion. Then, the corresponding findings were played side by side and evaluated by fourteen ultrasound experts using a point scale (5 points = very good; 1 point = insufficient).
Results: The HHUS devices achieved relatively good results in terms of both the B-scan quality assessment and the ability to answer the clinical question, regardless of the manufacturer. One of the tested HHUS devices even achieved a significantly ( p < 0.05) higher average points score in both the evaluation of B-scan quality and in the evaluation of clinical significance than the other devices. Regardless of the manufacturer, the HHUS devices performed best when determining the status/inferior vena cava volume and in the representation of ascites/free fluid.
Conclusion: In various clinical abdominal sonography questions, HHUS systems can reliably reproduce findings, and are-while bearing their limitations in mind-an acceptable alternative to conventional HEUS systems. Irrespective of this, the present study demonstrated relevant differences in the B-scan quality of HHUS devices from different manufacturers.