Autor: |
Cameron CJ, Fortune BC, Pretty CG, Hayes MP |
Jazyk: |
angličtina |
Zdroj: |
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2023 Jul; Vol. 2023, pp. 1-4. |
DOI: |
10.1109/EMBC40787.2023.10340097 |
Abstrakt: |
This paper presents a method for identifying parameter values for a double parallel resistor/constant-phase-element model of the electrode-skin interface for individual silver and silver/silver chloride electrodes. The impedance of each electrode was measured in five from 1 Hz-10 kHz. Phase features of these data were used to guide initial estimates for parameter values which were refined using a least squares algorithm. Resultant model impedances were compared with experimental data across a typical biosignal bandwidth (1 Hz-500 Hz). The method was effective in estimating component values in most datasets, and resulted in a mean relative RMS error of 7 % (σ = 8.3%) across the biosignal bandwidth.Clinical relevance- This work establishes a feature-based method for finding component parameter estimates for an electrode contact impedance model. |
Databáze: |
MEDLINE |
Externí odkaz: |
|