Minimising alarm pressure on a single room NICU through automated withdrawal of resolved alarms.

Autor: Broer SDL; Division of Neonatology, Department of Pediatrics, Willem-Alexander Children's Hospital, Leiden University Medical Center, Leiden, the Netherlands., Cramer SJE; Division of Neonatology, Department of Pediatrics, Willem-Alexander Children's Hospital, Leiden University Medical Center, Leiden, the Netherlands., Tan RNGB; Division of Neonatology, Department of Pediatrics, Willem-Alexander Children's Hospital, Leiden University Medical Center, Leiden, the Netherlands., Witlox RSGM; Division of Neonatology, Department of Pediatrics, Willem-Alexander Children's Hospital, Leiden University Medical Center, Leiden, the Netherlands.
Jazyk: angličtina
Zdroj: Acta paediatrica (Oslo, Norway : 1992) [Acta Paediatr] 2024 Feb; Vol. 113 (2), pp. 206-211. Date of Electronic Publication: 2023 Nov 15.
DOI: 10.1111/apa.17029
Abstrakt: Aim: In 2017, the Leiden University Medical Centre implemented a secondary alarm system using handheld devices to ensure accurate patient monitoring on the single room NICU. Initially, alarms remained active on the handheld devices until one of the caregivers in the alarm chain accepted the alarm. In 2020, a bidirectional communication protocol (BCP) was implemented, enabling automated withdrawal of resolved alarms. The aim of this study was to evaluate the effect of this implementation on the alarm duration and pressure.
Methods: Data of all alarms of the secondary alarm chain in the 90 days before and after the implementation were analysed and compared between both periods.
Results: Following the implementation of the BCP, 60% of the alarms were withdrawn before the designated nurse responded. Despite a significant higher total number of alarms, the median alarm duration decreased from 9 (7-14) to 6 (4-10) s, the acceptance rate of the designated nurse increased from 93% to 95% and the median time of alarm sounding per phone per hour significantly decreased from 71 (51-101) to 51 (35-69) s following implementation of the BCP.
Conclusion: This study showed that automated withdrawal of resolved alarms significantly reduces alarm duration and pressure on a NICU.
(© 2023 The Authors. Acta Paediatrica published by John Wiley & Sons Ltd on behalf of Foundation Acta Paediatrica.)