Sensing of concealed ventricular extrasystoles by pacemakers. Fifty years later.
Autor: | Barold SS; University of Rochester School of Medicine and Dentistry, Rochester, New York, USA., Kucher A; BIOTRONIK SE & Co. KG, Berlin, Germany. |
---|---|
Jazyk: | angličtina |
Zdroj: | Journal of cardiovascular electrophysiology [J Cardiovasc Electrophysiol] 2023 Dec; Vol. 34 (12), pp. 2607-2612. Date of Electronic Publication: 2023 Nov 14. |
DOI: | 10.1111/jce.16121 |
Abstrakt: | Whether a pacemaker can sense concealed ventricular extrasystoles still remains debatable since its occurrence was first proposed in 1972. It must remain a diagnosis of exclusion if it really exists. Isoelectric complexes and all the causes of oversensing especially discrete false signals generated by a defective pacemaker lead must be excluded before concealed ventricular extrasystoles can be postulated. (© 2023 Wiley Periodicals LLC.) |
Databáze: | MEDLINE |
Externí odkaz: |