Positional accuracy during the sequence of static computer-assisted implant surgery in three alveolar ridge morphologies: An in vitro study.

Autor: Raabe C; Department of Oral Surgery and Stomatology, School of Dental Medicine, University of Bern, Bern, Switzerland., Abou-Ayash S; Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland., Yilmaz B; Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland.; Department of Restorative, Preventive and Pediatric Dentistry, School of Dental Medicine, University of Bern, Bern, Switzerland.; Division of Restorative and Prosthetic Dentistry, The Ohio State University College of Dentistry, Columbus, Ohio, USA., Surbek FJ; Department of Oral Surgery and Stomatology, School of Dental Medicine, University of Bern, Bern, Switzerland., Chappuis V; Department of Oral Surgery and Stomatology, School of Dental Medicine, University of Bern, Bern, Switzerland., Couso-Queiruga E; Department of Oral Surgery and Stomatology, School of Dental Medicine, University of Bern, Bern, Switzerland.
Jazyk: angličtina
Zdroj: Journal of prosthodontics : official journal of the American College of Prosthodontists [J Prosthodont] 2023 Nov 13. Date of Electronic Publication: 2023 Nov 13.
DOI: 10.1111/jopr.13798
Abstrakt: Purpose: This in vitro study aimed to assess the positional accuracy during the sequence of static computer-assisted implant surgery (sCAIS) according to the anatomical characteristics of the alveolar ridge.
Material and Methods: Maxillary bone models with six single tooth gaps including clinical scenarios of healed alveolar ridge (HR), single-rooted (SRS), and three-rooted socket (TRS) morphologies were used in this study. Positional deviations during implant placement procedures were evaluated after the pilot osteotomy (PD), final osteotomy (FD), and implant placement with respect to the pre-planned implant position by using a software package. ANOVA and post hoc analyses were performed.
Results: A total of 90 implants were included in this study. Higher mean angular, crestal, and apical deviations were found after the PD and FD (3.5 ± 2.4°, 0.7 ± 0.3 mm, and 1.4 ± 0.8 mm versus 3.6 ±2.2°, 0.6 ± 0.3 mm, and 1.2 ± 0.6 mm) compared to IP (2.8 ± 1.6°, 0.7 ± 0.3 mm, and 1.2 ± 0.5 mm, p ≤ 0.004). Implants placed in TRS demonstrated higher mean angular, crestal, and apical deviations (4.0 ± 1.7°, 0.8 ± 0.3 mm, and 1.6 ± 0.5 mm) compared to implants placed in SRS (2.5 ± 1.2°, 0.7 ± 0.3 mm, and 1.1 ± 0.4 mm) or HR (2.0 ± 0.9°, 0.5 ± 0.3 mm, and 0.8 ± 0.4 mm, p < 0.001).
Conclusions: Positional deviations during sCAIS procedures are initiated with the first implant osteotomy and persist throughout the drilling sequence. However, deviations slightly decreased after implant placement. The alveolar ridge morphology is strongly associated with positional deviations. Higher deviations were observed in three-rooted and single-rooted sockets simulating an immediate approach compared to healed sites simulating a delayed protocol.
(© 2023 by the American College of Prosthodontists.)