Local SAR management strategies to use two-channel RF shimming for fetal MRI at 3 T.
Autor: | Yetisir F; Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Boston, Massachusetts, USA., Abaci Turk E; Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Boston, Massachusetts, USA.; Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, USA., Adalsteinsson E; Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.; Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA., Wald LL; Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.; Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA., Grant PE; Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Boston, Massachusetts, USA.; Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, USA.; Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA. |
---|---|
Jazyk: | angličtina |
Zdroj: | Magnetic resonance in medicine [Magn Reson Med] 2024 Mar; Vol. 91 (3), pp. 1165-1178. Date of Electronic Publication: 2023 Nov 06. |
DOI: | 10.1002/mrm.29913 |
Abstrakt: | Purpose: This study evaluates the imaging performance of two-channel RF-shimming for fetal MRI at 3 T using four different local specific absorption rate (SAR) management strategies. Methods: Due to the ambiguity of safe local SAR levels for fetal MRI, local SAR limits for RF shimming were determined based on either each individual's own SAR levels in standard imaging mode (CP mode) or the maximum SAR level observed across seven pregnant body models in CP mode. Local SAR was constrained either indirectly by further constraining the whole-body SAR (wbSAR) or directly by using subject-specific local SAR models. Each strategy was evaluated by the improvement of the transmit field efficiency (average |B Results: Constraining wbSAR when using RF shimming decreases B Conclusion: Two-channel RF-shimming can safely and significantly improve the transmit field inside the fetus when subject-specific models are used with local SAR limits based on maximum CP mode SAR levels in the pregnant population. (© 2023 International Society for Magnetic Resonance in Medicine.) |
Databáze: | MEDLINE |
Externí odkaz: |