Validation of mixed-reality surgical navigation for glenoid axis pin placement in shoulder arthroplasty using a cadaveric model.
Autor: | Sanchez-Sotelo J; Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN, USA. Electronic address: sanchezsotelo.joaquin@mayo.edu., Berhouet J; Department of Orthopedic Surgery, University of Tours, Tours, France., Chaoui J; Stryker, Lyon, France., Freehill MT; Department of Orthopedic Surgery, Stanford University, San Francisco, CA, USA., Collin P; Department of Orthopedic Surgery, American Hospital of Paris, Paris, France., Warner J; Department of Orthopedic Surgery, Harvard University, Boston, MA, USA., Walch G; Department of Orthopedic Surgery, Centre Orthopédique Santy, Lyon, France., Athwal GS; Department of Orthopedic Surgery, Roth McFarlane Hand & Upper Limb Centre, London, ON, Canada. |
---|---|
Jazyk: | angličtina |
Zdroj: | Journal of shoulder and elbow surgery [J Shoulder Elbow Surg] 2024 May; Vol. 33 (5), pp. 1177-1184. Date of Electronic Publication: 2023 Oct 26. |
DOI: | 10.1016/j.jse.2023.09.027 |
Abstrakt: | Background: Mixed reality may offer an alternative for computer-assisted navigation in shoulder arthroplasty. The purpose of this study was to determine the accuracy and precision of mixed-reality guidance for the placement of the glenoid axis pin in cadaver specimens. This step is essential for accurate glenoid placement in total shoulder arthroplasty. Methods: Fourteen cadaveric shoulders underwent simulated shoulder replacement surgery by 7 experienced shoulder surgeons. The surgeons exposed the cadavers through a deltopectoral approach and then used mixed-reality surgical navigation to insert a guide pin in a preplanned position and trajectory in the glenoid. The mixed-reality system used the Microsoft Hololens 2 headset, navigation software, dedicated instruments with fiducial marker cubes, and a securing pin. Computed tomography scans obtained before and after the procedure were used to plan the surgeries and determine the difference between the planned and executed values for the entry point, version, and inclination. One specimen had to be discarded from the analysis because the guide pin was removed accidentally before obtaining the postprocedure computed tomography scan. Results: Regarding the navigated entry point on the glenoid, the mean difference between planned and executed values was 1.7 ± 0.8 mm; this difference was 1.2 ± 0.6 mm in the superior-inferior direction and 0.9 ± 0.8 mm in the anterior-posterior direction. The maximum deviation from the entry point for all 13 specimens analyzed was 3.1 mm. Regarding version, the mean difference between planned and executed version values was 1.6° ± 1.2°, with a maximum deviation in version for all 13 specimens of 4.1°. Regarding inclination, the mean angular difference was 1.7° ± 1.5°, with a maximum deviation in inclination of 5°. Conclusions: The mixed-reality navigation system used in this study allowed surgeons to insert the glenoid guide pin on average within 2 mm from the planned entry point and within 2° of version and inclination. The navigated values did not exceed 3 mm or 5°, respectively, for any of the specimens analyzed. This approach may help surgeons more accurately place the definitive glenoid component. (Copyright © 2023 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Elsevier Inc. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |