Size and Shape Matter: Cell Viability of Preloaded Descemet Stripping Automated Endothelial Keratoplasty Grafts in Three Different Carriers.

Autor: Straiko MMW; VisionGift, Portland, OR., Sawant OB; Center for Vision and Eye Banking Research, Eversight, Cleveland, OH; and., Hubbs R; VisionGift, Portland, OR., Dye PK; VisionGift, Portland, OR., Tsering D; VisionGift, Portland, OR., Hicks N; Department of Clinical Operations, Eversight, Ann Arbor, MI., Odell K; VisionGift, Portland, OR., Ellison MS; VisionGift, Portland, OR., Titus MS; Department of Clinical Operations, Eversight, Ann Arbor, MI., Straiko MD; Devers Eye Institute, Portland, OR., Tran KD; VisionGift, Portland, OR.
Jazyk: angličtina
Zdroj: Cornea [Cornea] 2024 Jan 01; Vol. 43 (1), pp. 111-119. Date of Electronic Publication: 2023 Sep 22.
DOI: 10.1097/ICO.0000000000003385
Abstrakt: Purpose: The aim of this study was to examine endothelial cell loss (ECL) associated with preloaded Descemet stripping automated endothelial keratoplasty (DSAEK) grafts loaded into 3 carriers of different size and shape.
Methods: Thirty-six donor corneas were prepared for DSAEK and loaded into an EndoGlide Ultrathin (control) (2.0 mm × 3.5 mm lumen, 4.5 mm/4.9 mm incision for scleral tunnel/clear corneal insertion), Descemet membrane endothelial keratoplasty EndoGlide (experimental) (1.1 mm × 1.7 mm lumen, 2.65 mm incision), or round glass Jones tube (experimental) (1.8 mm lumen, 3.0 mm incision). Control grafts were stored for 6 days in Optisol-GS and experimental grafts stored for 24 hours in Life4C before analysis using Calcein AM staining. Grafts were imaged and ECL was analyzed by FIJI segmentation. The statistical significance of ECL was determined using 1-way ANOVA and Tukey post hoc analysis.
Results: There were no significant differences in donor characteristics for grafts in each cohort. ECL for grafts loaded into the EndoGlide Ultrathin was 10.3% ± 2.3% (graft thickness: 60-189 μm, n = 9). ECL for grafts loaded into the Descemet membrane endothelial keratoplasty EndoGlide was 22.2% ± 7.1% (graft thickness: 38-63 μm, n = 9). ECL for thin grafts (34-60 μm, n = 9) loaded into the Jones tube was 24.0% ± 5.0%. ECL for thick grafts (92-119 μm, n = 9) loaded into the Jones tube was 34.2% ± 6.1% ECL ( P < 0.001). Combined regression analysis revealed that graft thickness is directly correlated to ECL ( P < 0.01).
Conclusions: The size and shape of the carrier can influence the cell viability of preloaded DSAEK grafts. Compared with a larger carrier, smaller lumen carriers are associated with greater ECL. In smaller lumen carriers, ECL increases with increasing DSAEK graft thickness.
Competing Interests: The authors have no funding or conflicts of interest to disclose.
(Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.)