A simulation study to evaluate the effect of 2D antiscatter grid primary transmission on flat panel detector based CBCT image quality.

Autor: Eldib ME; Department of Radiation Oncology, University of Colorado School of Medicine, 1665 Aurora Court, Suite 1032, Mail stop F-706, Aurora, CO 80045, United States of America., Bayat F; Department of Radiation Oncology, University of Colorado School of Medicine, 1665 Aurora Court, Suite 1032, Mail stop F-706, Aurora, CO 80045, United States of America., Miften M; Department of Radiation Oncology, University of Colorado School of Medicine, 1665 Aurora Court, Suite 1032, Mail stop F-706, Aurora, CO 80045, United States of America., Altunbas C; Department of Radiation Oncology, University of Colorado School of Medicine, 1665 Aurora Court, Suite 1032, Mail stop F-706, Aurora, CO 80045, United States of America.
Jazyk: angličtina
Zdroj: Biomedical physics & engineering express [Biomed Phys Eng Express] 2023 Oct 04; Vol. 9 (6). Date of Electronic Publication: 2023 Oct 04.
DOI: 10.1088/2057-1976/acfb8a
Abstrakt: Purpose . Two-dimensional antiscatter grids' (2D-ASGs) septal shadows and their impact on primary transmission play a critical role in cone-beam computed tomography (CBCT) image noise and artifact characteristics. Therefore, a numerical simulation platform was developed to evaluate the effect of 2D-ASG's primary transmission on image quality, as a function of grid geometry and CBCT system properties. Methods . To study the effect of 2D-ASG's septal shadows on primary transmission and CBCT image quality, two new methods were introduced; one to simulate projection signal gradients in septal shadows, and the other to simulate septal shadow variations due to gantry flex. Signal gradients in septal shadows were simulated by generating a system point spread function that was directly extracted from projection images of 2D-ASG prototypes in experiments. Variations in septal shadows due to gantry flex were simulated by generating oversampled shadow profiles extracted from experiments. Subsequently, the effect of 2D-ASG's septal shadows on primary transmission and image quality was evaluated. Results. For an apparent septal thickness of 0.15 mm, the primary transmission of 2D-ASG varied between 72%-90% for grid pitches 1-3 mm. In low-contrast phantoms, the effect of 2D-ASG's radiopaque footprint on information loss was subtle. At high spatial frequencies, information loss manifested itself as undersampling artifacts, however, its impact on image quality is subtle when compared to quantum noise. Effects of additive electronic noise and gantry flex induced ring artifacts on image quality varied as a function of grid pitch and septal thickness. Such artifacts were substantially less in lower resolution images. Conclusion . The proposed simulation platform allowed successful evaluation of CBCT image quality variations as a function of 2D-ASG primary transmission properties and CBCT system characteristics. This platform can be potentially used for optimizing 2D-ASG design properties based on the imaging task and properties of the CBCT system.
(© 2023 IOP Publishing Ltd.)
Databáze: MEDLINE