Background illumination and automated perimetry.

Autor: Klewin KM, Radius RL
Jazyk: angličtina
Zdroj: Archives of ophthalmology (Chicago, Ill. : 1960) [Arch Ophthalmol] 1986 Mar; Vol. 104 (3), pp. 395-7.
DOI: 10.1001/archopht.1986.01050150095035
Abstrakt: Visual field function in the right and left eyes of 31 normal volunteers was evaluated with an automated projection perimeter (OCTOPUS). Serial visual field evaluations were repeated in these same eyes with neutral filters of increasing optical density. We compared the results of threshold determinations with the different neutral filters in place before the examined eye. Significant reduction in threshold sensitivity at several test spots throughout the central 30 degrees visual field was seen with neutral density filters of 0.5 log units or greater. The low level of background illumination of the OCTOPUS perimeter (4.0 apostilbs) may allow relatively minor reduction in light transmission by the ocular media to produce significant changes in the recorded level of threshold sensitivity during visual field evaluation.
Databáze: MEDLINE