Activity standardization of 99m Tc by digitizer-based 4πce(LS)-γ and 4πce(PC)-γ coincidence counting.
Autor: | Takács MP; Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, Germany. Electronic address: marcell.takacs@ptb.de., Kossert K; Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, Germany., Nähle O; Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, Germany. |
---|---|
Jazyk: | angličtina |
Zdroj: | Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2023 Oct; Vol. 200, pp. 110962. Date of Electronic Publication: 2023 Jul 26. |
DOI: | 10.1016/j.apradiso.2023.110962 |
Abstrakt: | An aqueous solution of 99m Tc was standardized by means of 4πce(LS)-γ and 4πce(PC)-γ coincidence counting. The activity concentration was obtained by efficiency extrapolation relying on four different variation techniques. In both setups, fast digitizers were used for data acquisition. Detected events were recorded in list mode format, and their coincidence relationship was analyzed in the subsequent offline analysis. The combined result was used to participate in an international comparison using the Transfer Instrument of the International Reference System. Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. (Copyright © 2023 Elsevier Ltd. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |