The lost balloon at midnight: a case report reveals the inevitability of heart team existence.
Autor: | Elbayomi M; Department of Cardiovascular Surgery, Friedrich-Alexander-University, Krankenhausstr. 12, 91054, Erlangen, Germany. mohamed.elbayomi@uk-erlangen.de., Weyand M; Department of Cardiovascular Surgery, Friedrich-Alexander-University, Krankenhausstr. 12, 91054, Erlangen, Germany., Nooh E; Department of Cardiovascular Surgery, Friedrich-Alexander-University, Krankenhausstr. 12, 91054, Erlangen, Germany., Harig F; Department of Cardiovascular Surgery, Friedrich-Alexander-University, Krankenhausstr. 12, 91054, Erlangen, Germany. |
---|---|
Jazyk: | angličtina |
Zdroj: | Journal of cardiothoracic surgery [J Cardiothorac Surg] 2023 Apr 06; Vol. 18 (1), pp. 102. Date of Electronic Publication: 2023 Apr 06. |
DOI: | 10.1186/s13019-023-02202-8 |
Abstrakt: | Background: Dislodgement of a coronary stent-balloon catheter during percutaneous coronary intervention (PCI) is rare but is a life-threatening complication. A 57- year-old male presented with a non-ST elevation myocardial infarction (NSTEMI). Coronary angiography revealed total thrombotic occlusion of the Right coronary artery (RCA). Following the balloon dilatation of the RCA and while trying to retrieve the balloon catheter, the balloon was dislodged from the catheter shaft and entrapped in the coronary vessel. Under cardiopulmonary bypass, with antegrade cardioplegic arrest, the balloon was extracted through a coronary arteriotomy. Right coronary revascularization was done with reversed saphenous vein graft (SVG). Discussion: Given the variety of equipment that can be retained in the coronary artery and the multitude of mechanisms by which it may be entrapped, there are no straightforward techniques applicable to all situations. Specific guidelines or recommendations on properly managing these potentially life-threatening complications do not exist. However, the most crucial issue in the management of these cases is the hemodynamic status of the patient as well as the coronary flow in the vessel with entrapped device or stent. In our case, the RCA was retrogradely perfused from the left coronary artery, which provided time to transfer the patient to cardiovascular surgical backup. (© 2023. The Author(s).) |
Databáze: | MEDLINE |
Externí odkaz: | |
Nepřihlášeným uživatelům se plný text nezobrazuje | K zobrazení výsledku je třeba se přihlásit. |