The outcomes of a low-cost, non-valved glaucoma drainage device using mitomycin-C: 1-year results.

Autor: Awad-Allah MAA; Ophthalmology Department, Faculty of Medicine, Ain Shams University, Cairo, Egypt. m.abd_elrazeik@yahoo.com., Mousa AS; Ophthalmology Department, Faculty of Medicine, Ain Shams University, Cairo, Egypt., Ashour DM; Ophthalmology Department, Faculty of Medicine, Ain Shams University, Cairo, Egypt.
Jazyk: angličtina
Zdroj: Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie [Graefes Arch Clin Exp Ophthalmol] 2023 Aug; Vol. 261 (8), pp. 2343-2349. Date of Electronic Publication: 2023 Mar 04.
DOI: 10.1007/s00417-023-06019-y
Abstrakt: Purpose: To evaluate the indications, outcomes, and complications of the usage of Aurolab Aqueous Drainage Implant (AADI) using mitomycin-C.
Methods: A retrospective case series of patients who underwent AADI placement using mitomycin-C between April 2018 and June 2020 at Ain Shams University Hospitals, Cairo, Egypt. The data was extracted from the records of the patients with a minimum of 1 year of follow-up. Complete success was defined as IOP ≥ 5 mmHg and ≤ 21 mmHg or reduction of IOP by ≥ 20% from baseline without antiglaucoma medications (AGMs). Qualified success was defined as reaching the same IOP range with the aid of AGM.
Results: A total of 50 eyes of 48 patients were included. Neovascular glaucoma represented the commonest indication (13 patients, 26%). The mean preoperative IOP was 34.0 ± 7.1 mmHg, with a median number of AGM of 3 (mean ± SD = 2.84 ± 1), while the mean IOP after 12 months was 14.3 ± 4 with a median number of AGM of 0. (mean ± SD = 0.52 ± 0.89) (p < 0.001). Complete success was achieved in 33 patients (66%). Qualified success was achieved in 14 patients (28%). Thirteen eyes (26%) had variable postoperative complications; none of them required explantation of the device or affected the visual acuity (except one patient).
Conclusion: AADI with using mitomycin-C and ripcord during the surgery is an effective and relatively safe method of control of IOP in refractory and advanced cases of glaucoma, with an overall success rate of 94%.
(© 2023. The Author(s).)
Databáze: MEDLINE