Beam output checks of a commercial high-field magnetic resonance-guided radiotherapy machine with its on-board megavoltage imager.
Autor: | Hilgers GC; Radiotherapiegroep, Arnhem/Deventer, The Netherlands., Ikink M; Radiotherapiegroep, Arnhem/Deventer, The Netherlands., Potters I; Radiotherapiegroep, Arnhem/Deventer, The Netherlands., Schuring D; Radiotherapiegroep, Arnhem/Deventer, The Netherlands. |
---|---|
Jazyk: | angličtina |
Zdroj: | Physics and imaging in radiation oncology [Phys Imaging Radiat Oncol] 2023 Jan 11; Vol. 25, pp. 100411. Date of Electronic Publication: 2023 Jan 11 (Print Publication: 2023). |
DOI: | 10.1016/j.phro.2023.100411 |
Abstrakt: | Beam output checks of a commercial high-field magnetic resonance-guided radiotherapy machine can be performed with its on-board megavoltage imager (MVI). This is a fast and efficient method, but only recommended for daily checks. The aim of our study was to show its suitability for weekly checks by investigating its long-term agreement with the golden standard: ionization chamber measurements in a water tank. For one year, the output deviations obtained with both methods were compared. The difference was 0.1 ± 0.3 (1SD) percentage point. This indicated an excellent agreement, and translated into a tolerance level of ± 2 %. Competing Interests: The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. (© 2023 The Author(s).) |
Databáze: | MEDLINE |
Externí odkaz: |