Characterisation of a split gradient coil design induced systemic imaging artefact on 0.35 T MR-linac systems.
Autor: | Marage L; Service de physique médicale, Centre Georges-François Leclerc, Dijon, France., Walker PM; Université de Bourgogne Franche-Comté, Laboratoire Imvia, Dijon, France., Boudet J; Service de physique médicale, Centre Georges-François Leclerc, Dijon, France., Fau P; Service de Radiothérapie, Institut Paoli-Calmettes, Marseille, France., Debuire P; Département de radiophysique, CRLC Val-d'Aurelle-Paul-Lamarque, Montpellier, France., Clausse E; Groupe Hospitalier Pitié-Salpêtrière, Assistance Publique - Hôpitaux de Paris, Service de Radiothérapie Oncologique, Paris, France., Petitfils A; Service de physique médicale, Centre Georges-François Leclerc, Dijon, France., Aubignac L; Service de physique médicale, Centre Georges-François Leclerc, Dijon, France., Rapacchi S; Université Aix-Marseille, CNRS, CRMBM, Marseille, France., Bessieres I; Service de physique médicale, Centre Georges-François Leclerc, Dijon, France. |
---|---|
Jazyk: | angličtina |
Zdroj: | Physics in medicine and biology [Phys Med Biol] 2022 Dec 29; Vol. 68 (1). Date of Electronic Publication: 2022 Dec 29. |
DOI: | 10.1088/1361-6560/aca876 |
Abstrakt: | Objective . The aim of this work was to highlight and characterize a systemic 'star-like' artefact inherent to the low field 0.35 T MRIdian MR-linac system, a magnetic resonance guided radiotherapy device. This artefact is induced by the original split gradients coils design. This design causes a surjection of the intensity gradient in Z (or head-feet) direction. This artefact appears on every sequence with phase encoding in the head-feet direction. Approach . Basic gradient echo sequence and clinical mandatory bSSFP sequence were used. Three setups using manufacturer provided QA phantoms were designed: two including the linearity control grid used for the characterisation and a third including two homogeneity control spheres dedicated to the artefact management in a more clinical like situation. The presence of the artefact was checked in four different MRidian sites. The tested parameters based on the literature were: phase encoding orientation, slab selectivity, excitation bandwidth (BW (© 2022 Institute of Physics and Engineering in Medicine.) |
Databáze: | MEDLINE |
Externí odkaz: |