Evaluation of data-driven respiratory gating for subcentimeter lesions using digital PET/CT system and three-axis motion phantom.
Autor: | Fukai S; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan.; Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, 323-1 Kamioki-machi, Maebashi, Gunma 371-0052, Japan., Daisaki H; Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, 323-1 Kamioki-machi, Maebashi, Gunma 371-0052, Japan.; Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8510, Japan., Shimada N; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan., Ishiyama M; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan., Umeda T; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan., Yamashita K; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan., Miyaji N; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan., Takiguchi T; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan., Kawakami H; APEX Medical, Inc., Kuramae Myouken-ya Building 5F, 3-17-4 Kuramae, Taito-ku, Tokyo 111-0051, Japan., Terauchi T; Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan. |
---|---|
Jazyk: | angličtina |
Zdroj: | Biomedical physics & engineering express [Biomed Phys Eng Express] 2022 Dec 16; Vol. 9 (1). Date of Electronic Publication: 2022 Dec 16. |
DOI: | 10.1088/2057-1976/aca90d |
Abstrakt: | Introduction. The application of data-driven respiratory gating (DDG) for subcentimeter lesions with respiratory movement remains poorly understood. Hence, this study aimed to clarify DDG application for subcentimeter lesions and the ability of digital Positron emission tomography/computed tomography (PET/CT) system combined with DDG to detect these lesions under three-axis respiration. Methods. Discovery MI PET/CT system and National Electrical Manufacturers Association (NEMA) body phantom with Micro Hollow Sphere (4, 5, 6, 8, 10, and 13 mm) were used. The NEMA phantom was filled with 18 F-FDG solutions of 42.4 and 5.3 kBq/ml for each hot sphere and background region. The 3.6 s cycles of three-axis respiratory motion were reproduced using the motion platform UniTraQ. The PET data acquisition was performed in stationary and respiratory-moving states. The data were reconstructed in three PET groups: stationary (NM-PET), no gating with respiratory movement (NG-PET), and DDG gating with respiratory movement (DDG-PET) groups. For image quality, percent contrast (Q (© 2022 IOP Publishing Ltd.) |
Databáze: | MEDLINE |
Externí odkaz: |