Toward FBG-Sensorized Needle Shape Detection in Real Tissue Insertions.

Autor: Kim MJ, Lezcano DA, Kim JS, Iordachita II
Jazyk: angličtina
Zdroj: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2022 Jul; Vol. 2022, pp. 4397-4401.
DOI: 10.1109/EMBC48229.2022.9871525
Abstrakt: The determination of flexible needle shape during insertion is critical for planning and validation in minimally invasive surgical percutaneous procedures. In this paper, we validate a needle shape-sensing method using fiber Bragg grating (FBG) sensors over sequential needle insertion lengths in gel phantom and real tissue. Experiments on a four-active area, FBG-sensorized needle were performed in both isotropic simulated tissue and inhomogeneous animal tissue with computed tomography (CT) as the ground truth of the needle shape. The results show that the needle shape obtained from the FBG sensors has an overall consistent accuracy in real tissue in comparison to the phantom gel. The results validate a viable 3D needle shape-sensing model and reconstruction method over various insertion depths in comparison to the needle shapes determined from CT in both gel phantom and real tissue.
Databáze: MEDLINE