In Vivo Determination of the Human Corneal Elastic Modulus Using Vibrational Optical Coherence Tomography.
Autor: | Crespo MA; Cornea Service, Wills Eye Hospital, Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA, USA., Jimenez HJ; Vickie and Jack Farber Vision Research Center, Wills Eye Hospital, Philadelphia, PA, USA., Deshmukh T; OptoVibronex, LLC, Bethlehem, PA, USA., Pulido JS; Vickie and Jack Farber Vision Research Center, Wills Eye Hospital, Philadelphia, PA, USA., Saad AS; Cornea Service, Wills Eye Hospital, Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA, USA., Silver FH; OptoVibronex, LLC, Bethlehem, PA, USA.; Department of Pathology and Laboratory Medicine, Robert Wood Johnson Medical School, Piscataway, NJ, USA., Benedetto DA; Center for Advanced Eye Care, Vero Beach, FL, USA., Rapuano CJ; Cornea Service, Wills Eye Hospital, Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA, USA., Syed ZA; Cornea Service, Wills Eye Hospital, Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA, USA. |
---|---|
Jazyk: | angličtina |
Zdroj: | Translational vision science & technology [Transl Vis Sci Technol] 2022 Jul 08; Vol. 11 (7), pp. 11. |
DOI: | 10.1167/tvst.11.7.11 |
Abstrakt: | Purpose: To determine the in vivo elastic modulus of the human cornea using vibrational optical coherence tomography (VOCT). Methods: Vibrational analysis coupled with optical coherence tomography (OCT) was used to obtain the resonant frequency (RF) and elastic modulus of corneal structural components. VOCT corneal thickness values were measured using OCT images and correlated with corneal thickness determined with Pentacam (Oculus, Wetzlar, Germany). Moduli were obtained at two locations: central cornea (CC) and inferior cornea (IC). Measurements were obtained with and without anesthetic eye drops to assess their effect on the modulus measurements. Results: VOCT thickness values correlated positively (R2 = 0.97) and linearly (y = 1.039x-16.89) with those of Pentacam. Five RF peaks (1-5) were present, although their presence was variable across eyes. The RF for peaks 1 to 5 in the CC and IC ranged from 73.5 ± 4.9 to 239 ± 3 Hz and 72.1 ± 6.3 to 238 ± 4 Hz, respectively. CC and IC moduli for peaks 1 to 5 ranged from 1.023 ± 0.104 to 6.87 ± 0.33 MPa and 0.98 ± 0.15 to 6.52 ± 0.79 MPa, respectively. Topical anesthesia did not significantly alter the modulus (P > 0.05 for all), except for peak 2 in the CC (P < 0.05). Conclusions: This pilot study demonstrates the utility of VOCT as an in vivo, noninvasive technology to measure the elastic modulus in human corneas. The structural origin of these moduli is hypothesized based on previous reports, and further analyses are necessary for confirmation. Translational Relevance: This work presents VOCT as a novel approach to assess the in vivo elastic modulus of the cornea, an indicator of corneal structural integrity and health. |
Databáze: | MEDLINE |
Externí odkaz: |