Evaluation of the DiaSorin LIAISON SARS-CoV-2 antigen assay on nasopharyngeal swabs in two different SARS-CoV-2 pandemic waves in Switzerland: The impact of the Omicron variant on its performance.

Autor: Uster S; Institute for Infectious Diseases, University of Bern, Friedbühlstrasse 51, 3001 Bern, Switzerland., Topalli Z; Institute for Infectious Diseases, University of Bern, Friedbühlstrasse 51, 3001 Bern, Switzerland., Sasse T; Institute for Infectious Diseases, University of Bern, Friedbühlstrasse 51, 3001 Bern, Switzerland., Suter-Riniker F; Institute for Infectious Diseases, University of Bern, Friedbühlstrasse 51, 3001 Bern, Switzerland., Barbani MT; Institute for Infectious Diseases, University of Bern, Friedbühlstrasse 51, 3001 Bern, Switzerland.
Jazyk: angličtina
Zdroj: Journal of clinical virology plus [J Clin Virol Plus] 2022 Aug; Vol. 2 (3), pp. 100095. Date of Electronic Publication: 2022 Jun 26.
DOI: 10.1016/j.jcvp.2022.100095
Abstrakt: Background: SARS-CoV-2 antigen tests reliably detect individuals with high viral loads and provide an efficient diagnostic tool to manage the current SARS-CoV-2 pandemic . However, mutations in SARS-CoV-2 variants of concerns that appeared after validation of most antigen tests might impact their diagnostic performance.
Objectives: To assess the impact of the Omicron variant on the performance of the DiaSorin LIAISON SARS-CoV-2 antigen test, we evaluated its sensitivity and specificity on nasopharyngeal swabs (NPS) compared to rRT-PCR in the second and the Omicron pandemic wave in Switzerland.
Study Design: A random selection of NPS from patients undergoing SARS-CoV-2 diagnostics by rRT-PCR were collected during the second and the Omicron pandemic wave and further analyzed by the LIAISON antigen test. Sensitivity and specificity compared to rRT-PCR were calculated.
Results: Test performance did not change in the two investigated periods. The overall sensitivity of 75.8% in the second and 76.5% in the Omicron wave increased to 87.1% and 88.4%, excluding samples with rRT-PCR Ct-value >30. By lowering the cut-off from 200 TCID 50 /ml to 62 TCID 50 /ml to discriminate between negative and positive samples using a ROC-curve, the sensitivity resulted in 88.8% for the second and 93.3% for the Omicron pandemic wave. The specificity of the LIAISON antigen test was 100% in both collectives.
Conclusion: Omicron variant does not seem to affect the performance of the LIAISON antigen test. The WHO recommended sensitivity of ≥80% for antigen testing was fulfilled during both pandemic periods in samples with Ct-value <30 or by optimizing the assay cut-off.