Data associated with a tellurium absorption atlas 19,000 - 24,000 cm -1 .

Autor: Ross AJ; Institut Lumière Matière, Université Claude Bernard Lyon 1, CNRS, Université de Lyon, Villeurbanne F-69622, France., Cardon JM; Ideal Vacuum Products, 5910 Midway Park Blvd NE, Albuquerque NM 87109, United States of America.
Jazyk: angličtina
Zdroj: Data in brief [Data Brief] 2022 Mar 09; Vol. 42, pp. 108038. Date of Electronic Publication: 2022 Mar 09 (Print Publication: 2022).
DOI: 10.1016/j.dib.2022.108038
Abstrakt: Transmittance and absorbance data in the form of a spectral atlas have been obtained from Fourier transform spectra recorded on a commercial (Bomem DA3) instrument, operated with an external white light source injected through the instrument's `emission' port. The sample was a sealed, evacuated cell containing a small quantity of 130 Te. This cell was placed in a ceramic furnace maintained close to 620 °C, with tellurium vapor pressures 8-11 Torr. The data are intended as an aid to spectral calibration in the blue-green to violet region. The data relate to work published in J. Mol. Spectrosc. 384 111,589 (2022).
Competing Interests: The authors declare the following financial interests/personal relationships which may be considered as potential competing interests. Although there is no obvious competing interest, we indicate that Joseph M. Cardon is an employee of Ideal Vacuum Products, a company that commercializes the tellurium reference cell and temperature-controlled furnace cited in the paper.
(© 2022 The Authors.)
Databáze: MEDLINE