Anatomical and Technical Considerations of Robot-Assisted Cervical Pedicle Screw Placement: A Cadaveric Study.
Autor: | Mao JZ; Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA.; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia PA, USA., Soliman MAR; Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA.; Department of Neurosurgery, Faculty of Medicine, Cairo University, Giza, Egypt., Karamian BA; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia PA, USA., Khan A; Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA., Fritz AG; Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA., Avasthi N; Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA., DiMaria S; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia PA, USA., Levy BR; George Washington University, School of Medicine and Health Sciences, Washington DC, USA., O'Connor TE; Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA., Schroeder G; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia PA, USA., Pollina J; Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA., Vaccaro AR; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia PA, USA., Mullin JP; Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo NY, USA. |
---|---|
Jazyk: | angličtina |
Zdroj: | Global spine journal [Global Spine J] 2023 Sep; Vol. 13 (7), pp. 1992-2000. Date of Electronic Publication: 2022 Feb 23. |
DOI: | 10.1177/21925682211068410 |
Abstrakt: | Study Design: Cadaver study. Objectives: Assess the feasibility of robot-assisted cervical pedicle screw (RA-CPS) placement and understand the anatomical considerations of this technique. Methods: Four cadaver specimens free from bony pathology were acquired. Anatomical considerations, such as pedicle width (PW) and height (PH), transverse pedicle angle (TPA), and maximal screw length (MSL), were recorded from preoperative computational tomography (CT) scans. Intraoperative cone-beam CT was acquired and registered to the robotic system. After cervical levels were segmented, screw sizes and trajectories were planned, and RA-CPS were placed. Accuracy was assessed using Gertzbein and Robbin's classification on postoperative CT scans. Results: Thirty-five RA-CPS were placed. Major breaches (≥Grade C) occurred in 28.57% screws. Grade A or B accuracy was found in 71.43% of screws, with the most common direction of breach being medial (81.3%). The greatest proportion of breach per level occurred in the upper subaxial levels, (C3:71.4%, C4 66.6%, C5:50%) which had the smallest PW (C3: 4.34 ± .96 mm, C4: 4.48 ± .60, C5: 5.76 ± 1.11). PH was greatest at C2 (8.14 ± 1.89 mm) and ranged subaxial from 6.36 mm (C3) to 7.48 mm (C7). The mean PW was 5.37 mm and increased caudally from 4.34 mm (C3) to 6.31 mm (C7). The mean TPA was 39.9° and decreased moving caudally 46.9°) to C7 (34.4°). The MSL was 37.1 mm and increased from C2 (26.3 mm) to C7 (41.0 mm). Conclusion: RA-CPS has the potential to be feasible, but technological and instrument modifications are necessary to increase the accuracy in the cervical region. |
Databáze: | MEDLINE |
Externí odkaz: |