Novel Cuffless Blood Pressure Estimation Method Using a Bayesian Hierarchical Model.

Autor: He S, Dajani HR, Bolic M
Jazyk: angličtina
Zdroj: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2021 Nov; Vol. 2021, pp. 898-901.
DOI: 10.1109/EMBC46164.2021.9629594
Abstrakt: Continuous blood pressure (BP) monitoring is important for the prevention and early diagnosis of cardiovascular diseases. Cuffless BP estimation using pulse arrival time (PAT) via a mathematical model which enables continuous BP measurement has recently become a popular research topic. In this study, simultaneous biomedical signals from ten healthy subjects were acquired by electrocardiogram (ECG) and photoplethysmogram (PPG) sensors and the continuous reference BP data were collected by a cuff-based Finometer PRO BP monitor. A hierarchical model was applied to estimate the parameters of a nonlinear model which in turn is used to estimate systolic blood pressure (SBP) using PAT with few calibration measurements. The mean absolute difference (MAD) between the estimated SBP and reference SBP is 4.35±1.43 mmHg using the proposed hierarchical model with three calibration measurements and is 4.36±1.17 mmHg with a single calibration measurement.
Databáze: MEDLINE