Current status of the recommended half-lives for the radionuclides used as the standards of photon emitters.
Autor: | Kuzmenko NK; V.G. Khlopin Radium Institute, 28 Second Murinsky Ave., St. Petersburg, 194021, Russia. Electronic address: kuzmenko@nk9433.spb.edu. |
---|---|
Jazyk: | angličtina |
Zdroj: | Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2022 Feb; Vol. 180, pp. 110039. Date of Electronic Publication: 2021 Nov 24. |
DOI: | 10.1016/j.apradiso.2021.110039 |
Abstrakt: | This paper presents an update of the recommended half-lives published in X- and gamma-ray standards for detector calibration (IAEA, 1991 and 2007) performed as part of the Coordinated Research Project under the auspices of the IAEA. The half-lives of 61 radionuclides were updated during the years 2018-2020 using a well-established methodology to select the dataset. The recommended half-lives were obtained using weighted or unweighted sample means and the consistency of the results was investigated and discussed based on the chi-square test. (Copyright © 2021 Elsevier Ltd. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |