Transient-mode photothermal optical coherence tomography.

Autor: Salimi MH, Villiger M, Tabatabaei N
Jazyk: angličtina
Zdroj: Optics letters [Opt Lett] 2021 Nov 15; Vol. 46 (22), pp. 5703-5706.
DOI: 10.1364/OL.443987
Abstrakt: Photothermal optical coherence tomography (PT-OCT) is an emerging extension of OCT, which forms images based on both scattering and absorption of light. The speed of PT-OCT, however, has been limited by the necessity for lock-in detection with extensive temporal sampling of the sample's PT response. Here, we demonstrate transient-mode PT-OCT (TM-PT-OCT), which increases the effective A-line rate by orders of magnitude from 10-100 Hz to 1.5-7.5 kHz, by interrogating the sample's transient thermal response to a single diode laser pulse. Functional imaging of moving samples with TM-PT-OCT at video rate is also presented. This significant improvement in imaging speed is expected to open the door for downstream integration of PT-OCT in clinical systems for in vivo imaging.
Databáze: MEDLINE