Do current Philips ultrasound systems exceed the recommended safety limits during routine prenatal ultrasounds?

Autor: Ong QH; Department of Ultrasound, Radiology Waikato Hospital Hamilton New Zealand., Necas M; Department of Ultrasound, Radiology Waikato Hospital Hamilton New Zealand., Lissington K; Hamilton Radiology Hamilton New Zealand.
Jazyk: angličtina
Zdroj: Australasian journal of ultrasound in medicine [Australas J Ultrasound Med] 2019 Jun 27; Vol. 22 (4), pp. 265-272. Date of Electronic Publication: 2019 Jun 27 (Print Publication: 2019).
DOI: 10.1002/ajum.12169
Abstrakt: Background: Biological safety of ultrasound is a complex and nuanced subject that is poorly understood by ultrasound users. Little is known about the acoustic output and thermal index levels during the routine use of modern ultrasound machines in prenatal scanning.
Methodology: This study was a retrospective review of thermal index (TI) values encountered during 300 prenatal ultrasound examinations (100 in each trimester) performed on any one of 13 Philips Epiq 7 or Epiq 5 systems, representing approximately 106.5 h of real-time scanning. The TI levels were compared to three international guidelines on the biological safety of ultrasound.
Results: The routine use of current Philips systems was associated with low TI levels. Of the 300 examinations reviewed, virtually all were compliant with the BMUS and Nelson safety guideline. Whether the examination was compliant with the WFUMB guideline is open to interpretation. The highest level of TI encountered was 1.1. In no instance did the TI level incur into the 'not recommended' range or into a range where specific user action was required to reduce the TI within 1 min. The most frequent action associated with TI > 0.7 was the use of M-mode to document the fetal heart rate. In the four instances where TI peaked at 1.1, 3 were associated with the use of M-mode and one with B-mode. Spectral Doppler was not implicated in high TI levels. These results are surprising and open up a range of opportunities for future study.
(© 2019 Australasian Society for Ultrasound in Medicine.)