An experimental pleural drainage device in hypertensive pneumothorax.

Autor: Petta BFV; Fellow Master degree. Postgraduate Program in Management, Technology and Innovation in Emergency and Urgency - Department of Medicine - Universidade Estadual de Maringá - Maringa (PR), Brazil., Cazanti RF; Fellow Master degree. Postgraduate Program in Management, Technology and Innovation in Emergency and Urgency - Department of Medicine - Universidade Estadual de Maringá - Maringa (PR), Brazil., Fontes CER; Post-Doctorate. University of Paris XI Paul Bousse Hospital and Department of Medicine - Universidade Estadual de Maringá - Maringa (PR), Brazil.
Jazyk: angličtina
Zdroj: Acta cirurgica brasileira [Acta Cir Bras] 2021 Aug 23; Vol. 36 (7), pp. e360708. Date of Electronic Publication: 2021 Aug 23 (Print Publication: 2021).
DOI: 10.1590/ACB360708
Abstrakt: Purpose: To develop a specific device for pleural drainage in hypertensive pneumothorax.
Methods: The prototype was modeled from the free version of a 3D modeling application, printed on a 3D printer using ABS® plastic material, and tested in a pleural drainage simulator.
Results: Pleural drainage in the simulator using the prototype was feasible and reproducible.
Conclusions: While the prototype is functional in the simulator, it requires improvement and refinement for use in humans.