Evaluation of Acuros® XB accuracy for static small fields dose calculations based on the IAEA/AAPM TRS-483 recommendation.

Autor: Lárraga-Gutiérrez JM; Lab. de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Insurgentes sur 3877, La Fama, Tlalpan 14269, CMDX, Mexico. Electronic address: jlarraga@innn.edu.mx., García-Garduño OA; Lab. de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Insurgentes sur 3877, La Fama, Tlalpan 14269, CMDX, Mexico., Herrera-González JA; Unidad de Radioneurocirugía, Instituto Nacional de Neurología y Neurocirugía, Insurgentes sur 3877, La Fama, Tlalpan 14269, CMDX, Mexico; Depto. de Biofísica, Instituto Nacional de Cancerología, San Fernando 22, Belisario Domínguez Secc 16, Tlalpan 14080, CDMX, Mexico., Galván de la Cruz OO; Unidad de Radioneurocirugía, Instituto Nacional de Neurología y Neurocirugía, Insurgentes sur 3877, La Fama, Tlalpan 14269, CMDX, Mexico.
Jazyk: angličtina
Zdroj: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2021 Sep; Vol. 89, pp. 140-146. Date of Electronic Publication: 2021 Aug 05.
DOI: 10.1016/j.ejmp.2021.06.021
Abstrakt: Purpose: Evaluate Acuros® XB dose calculation accuracy following TRS-483 recommendations in small static fields for flattened and un-flattened 6 MV X-ray beams.
Methods: Field output factors were measured following TRS-483 recommendations using four radiation detectors. Two sets of field output factors were measured. One set was used to configure the beam model into Acuros® XB down to a jaw-defined field size of 1.0 cm × 1.0 cm. The second set was used to evaluate the differences between calculated and measured field output factors for MLC-fields down to a field size of 0.5 cm × 0.5 cm.
Results: Acuros® XB showed an accuracy within 1.5% down to an MLC-field of 1.0 cm × 1.0 cm, for a focal spot size of 1.0 and 0.0 mm in the cross and in-plane directions. For an MLC-field of 0.5 cm × 0.5 cm, an agreement was found within 3% between calculated and measured field output factors. These results were addressed by optimizing the focal spot size to minimize the differences between calculated and measured dose profiles.
Conclusions: By optimizing the focal spot size, Acuros® XB showed an acceptable agreement within 3% down to an MLC-field of 0.5 cm × 0.5 cm. The results of this work suggest that if static and modulated delivery of very small targets is planned, then a field output factor table down to a field size of 1.0 cm is required in the beam configuration model.
(Copyright © 2021 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.)