Evaluation of the ARKRAY HA-8190V instrument for HbA 1c .
Autor: | van der Hagen EAE; European Reference Laboratory for Glycohemoglobin, Queen Beatrix Hospital, Winterswijk, The Netherlands.; Department of Clinical Chemistry, Queen Beatrix Hospital, Winterswijk, The Netherlands., Leppink S; Department of Clinical Chemistry, Queen Beatrix Hospital, Winterswijk, The Netherlands., Bokkers K; Department of Clinical Chemistry, Queen Beatrix Hospital, Winterswijk, The Netherlands., Siebelder C; European Reference Laboratory for Glycohemoglobin, Queen Beatrix Hospital, Winterswijk, The Netherlands.; Department of Clinical Chemistry, Queen Beatrix Hospital, Winterswijk, The Netherlands., Weykamp CW; European Reference Laboratory for Glycohemoglobin, Queen Beatrix Hospital, Winterswijk, The Netherlands.; Department of Clinical Chemistry, Queen Beatrix Hospital, Winterswijk, The Netherlands. |
---|---|
Jazyk: | angličtina |
Zdroj: | Clinical chemistry and laboratory medicine [Clin Chem Lab Med] 2020 Dec 24; Vol. 59 (5), pp. 965-970. Date of Electronic Publication: 2020 Dec 24 (Print Publication: 2021). |
DOI: | 10.1515/cclm-2020-1300 |
Abstrakt: | Objectives: Hemoglobin A Methods: The Clinical and Laboratory Standards Institute (CLSI) protocols EP-5, EP-9 and EP-10 were applied to investigate imprecision, bias and linearity. In addition potential interferences, performance in External Quality Assessment (EQA) and performance against the HA-8180V instrument in 220 clinical samples was evaluated. Results: The HA-8190V demonstrates a CV of ≤0.8% in IFCC SI units (≤0.6% National Glycohemoglobin Standardization Program [NGSP]) at 34 and 102 mmol/mol levels (5.3 and 11.5% NGSP) and a bias of -0.1 mmol/mol (-0.01% NGSP) at a concentration of 50 mmol/mol (6.7% NGSP), but with a significant slope as compared to target values. This results in a bias of -1.0 and 0.9 mmol/mol (-2.0 and 0.9% NGSP) at the 30 and 70 mmol/mol (4.9 and 8.6% NGSP) concentration level. Simulation of participation in the IFCC certification programme results in a Silver score (bias -0.1 mmol/mol, CV 1.1%). Interference in the presence of the most important Hb variants (AS, AC, AE, AD) and elevated HbA Conclusions: Analytical performance of the HA-8190V is very good, especially with respect to precision and HbA (© 2020 Walter de Gruyter GmbH, Berlin/Boston.) |
Databáze: | MEDLINE |
Externí odkaz: |