Autor: |
Yousefi T, Dabbaghian A, Kassiri H |
Jazyk: |
angličtina |
Zdroj: |
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2020 Jul; Vol. 2020, pp. 4479-4482. |
DOI: |
10.1109/EMBC44109.2020.9176671 |
Abstrakt: |
Motion artifacts are arguably the most important issue in the development of wearable ambulatory EEG devices. Designing circuits and systems capable of high-quality EEG recording regardless of these artifacts requires a clear understanding of how the electrode-skin interface is affected by physical motions. In this work, first, we report statistically-significant experimental characterization results of electrodeskin interface impedance for dry contact and non-contact electrodes in the presence of various motions. This leads to a model describing the motion-induced electrode-skin interface impedance variations for these electrodes. Next, a critical review of the possible analog front-end circuits for surface EEG recording is presented, followed by theoretical circuit analysis discussing the effect of electrode movements on the operation of these circuits. Inspired by the developed model and the analytical review, a novel front-end architecture capable of extracting motion from the EEG signal during the amplification stage is presented and experimentally characterized. |
Databáze: |
MEDLINE |
Externí odkaz: |
|