Is multileaf collimator tracking or gating a better intrafraction motion adaptation strategy? An analysis of the TROG 15.01 stereotactic prostate ablative radiotherapy with KIM (SPARK) trial.

Autor: Hewson EA; ACRF Image X Institute, University of Sydney Medical School, Australia. Electronic address: emily.hewson@sydney.edu.au., Nguyen DT; ACRF Image X Institute, University of Sydney Medical School, Australia; School of Biomedical Engineering, University of Technology Sydney., O'Brien R; ACRF Image X Institute, University of Sydney Medical School, Australia., Poulsen PR; Department of Oncology, Aarhus University Hospital, Denmark., Booth JT; Northern Sydney Cancer Centre, Royal North Shore Hospital, Australia; School of Physics, University of Sydney, Australia., Greer P; Calvary Mater Newcastle, Waratah, Australia., Eade T; Northern Sydney Cancer Centre, Royal North Shore Hospital, Australia., Kneebone A; Northern Sydney Cancer Centre, Royal North Shore Hospital, Australia., Hruby G; Northern Sydney Cancer Centre, Royal North Shore Hospital, Australia., Moodie T; Crown Princess Mary Cancer Centre, Sydney, Australia., Hayden AJ; Crown Princess Mary Cancer Centre, Sydney, Australia., Turner SL; Crown Princess Mary Cancer Centre, Sydney, Australia., Hardcastle N; Peter MacCallum Cancer Centre, Melbourne, Australia., Siva S; Peter MacCallum Cancer Centre, Melbourne, Australia., Tai KH; Peter MacCallum Cancer Centre, Melbourne, Australia., Martin J; Calvary Mater Newcastle, Waratah, Australia., Keall PJ; ACRF Image X Institute, University of Sydney Medical School, Australia.
Jazyk: angličtina
Zdroj: Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology [Radiother Oncol] 2020 Oct; Vol. 151, pp. 234-241. Date of Electronic Publication: 2020 Aug 20.
DOI: 10.1016/j.radonc.2020.08.010
Abstrakt: Purpose: Stereotactic Ablative Radiotherapy (SABR) has recently emerged as a favourable treatment option for prostate cancer patients. With higher doses delivered over fewer fractions, motion adaptation is a requirement for accurate delivery of SABR. This study compared the efficacy of multileaf collimator (MLC) tracking vs. gating as a real-time motion adaptation strategy for prostate SABR patients enrolled in a clinical trial.
Methods: Forty-four prostate cancer patients treated over five fractions in the TROG 15.01 SPARK trial were analysed in this study. Forty-nine fractions were treated using MLC tracking and 166 fractions were treated using beam gating and couch shifts. A time-resolved motion-encoded dose reconstruction method was used to evaluate the dose delivered using each motion adaptation strategy and compared to an estimation of what would have been delivered with no motion adaptation strategy implemented.
Results: MLC tracking and gating both delivered doses closer to the plan compared to when no motion adaptation strategy was used. Differences between MLC tracking and gating were small with differences in the mean discrepancy from the plan of -0.3% (CTV D 98% ), 1.4% (CTV D 2% ), 0.4% (PTV D 95% ), 0.2% (rectum V 30Gy ) and 0.0% (bladder V 30Gy ). On average, 0.5 couch shifts were required per gated fractions with a mean interruption duration of 1.8 ± 2.6 min per fraction treated using gating.
Conclusion: Both MLC tracking and gating were effective strategies at improving the accuracy of the dose delivered to the target and organs at risk. While dosimetric performance was comparable, gating resulted in interruptions to treatment.
Clinical Trial Registration Number: NCT02397317.