Reflection Versus Rebreathing for Administration of Sevoflurane During Minor Gynecological Surgery.

Autor: Kermad A; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany., Appenzeller M; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany., Morinello E; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany., Schneider SO; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany., Kleinschmidt S; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany., O'Gara B; Department of Anesthesiology, Beth Israel Deaconess Medical Center, Boston, Massachusetts., Volk T; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany., Meiser A; From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg/Saar, Germany.
Jazyk: angličtina
Zdroj: Anesthesia and analgesia [Anesth Analg] 2021 Apr 01; Vol. 132 (4), pp. 1042-1050.
DOI: 10.1213/ANE.0000000000005074
Abstrakt: Background: Contemporary anesthetic circle systems, when used at low fresh gas flows (FGF) to allow rebreathing of anesthetic, lack the ability for rapid dose titration. The small-scale anesthetic reflection device Anaesthetic Conserving Device (50mL Version; AnaConDa-S) permits administration of volatile anesthetics with high-flow ventilators. We compared washin, washout, and sevoflurane consumption using AnaConDa-S versus a circle system with low and minimal FGF.
Methods: Forty patients undergoing breast surgery were randomized to receive 0.5 minimal alveolar concentration (MAC) sevoflurane with AnaConDa-S (21 patients, reflection group) or with a circle system (low flow: FGF = 0.2 minute ventilation [V'E], 9 patients; or minimal flow: 0.1 V'E, 10 patients). In the reflection group, syringe pump boluses were given for priming and washin; to simulate an open system, the FGF of the anesthesia ventilator was set to 18 L·min-1 with the soda lime removed. In the other groups, the FGF was increased for washin (1 V'E for 8 minutes) and washout (3 V'E). For all patients, tidal volume was 7 mL·kg-1 and the respiratory rate adjusted to ensure normoventilation. Analgesia was attained with remifentanil 0.3 µg·kg-1·min-1. Sevoflurane consumption was compared between the reflection group and the low- and minimal-flow groups, respectively, using a post hoc test (Fisher Least Significant Difference). To compare washin and washout (half-life), the low- and minimal-flow groups were combined.
Results: Sevoflurane consumption was reduced in the reflection group (9.4 ± 2.0 vs 15.0 ± 3.5 [low flow, P < .001] vs 11.6 ± 2.3 mL·MAC h-1 [minimal flow, P = .02]); washin (33 ± 15 vs 49 ± 12 seconds, P = .001) and washout (28 ± 15 vs 55 ± 19 seconds, P < .001) times were also significantly shorter.
Conclusions: In this clinical setting with short procedures, low anesthetic requirements, and low tidal volumes, AnaConDa-S decreased anesthetic consumption, washin, and washout times compared to a circle system.