Autor: |
Slagowski JM; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, United States of America., Ding Y, Aima M, Wen Z, Fuller CD, Chung C, Debnam JM, Hwang KP, Kadbi M, Szklaruk J, Wang J |
Jazyk: |
angličtina |
Zdroj: |
Physics in medicine and biology [Phys Med Biol] 2020 Sep 28; Vol. 65 (19), pp. 195008. Date of Electronic Publication: 2020 Sep 28. |
DOI: |
10.1088/1361-6560/ab9c64 |
Abstrakt: |
Magnetic resonance imaging (MRI) offers outstanding soft tissue contrast that may reduce uncertainties in target and organ-at-risk delineation and enable online adaptive image-guided treatment. Spatial distortions resulting from non-linearities in the gradient fields and non-uniformity in the main magnetic field must be accounted for across the imaging field-of-view to prevent systematic errors during treatment delivery. This work presents a modular phantom and software application to characterize geometric distortion (GD) within the large field-of-view MRI images required for radiation therapy simulation. The modular phantom is assembled from a series of rectangular foam blocks containing high-contrast fiducial markers in a known configuration. The modular phantom design facilitates transportation of the phantom between different MR scanners and MR-guided linear accelerators and allows the phantom to be adapted to fit different sized bores or coils. The phantom was evaluated using a 1.5 T MR-guided linear accelerator (MR-Linac) and 1.5 T and 3.0 T diagnostic scanners. Performance was assessed by varying acquisition parameters to induce image distortions in a known manner. Imaging was performed using T1 and T2 weighted pulse sequences with 2D and 3D distortion correction algorithms and the receiver bandwidth (BW) varied as 250-815 Hz pixel -1 . Phantom set-up reproducibility was evaluated across independent set-ups. The software was validated by comparison with a non-modular phantom. Average geometric distortion was 0.94 ± 0.58 mm for the MR-Linac, 0.90 ± 0.53 mm for the 1.5 T scanner, and 1.15 ± 0.62 mm for the 3.0 T scanner, for a 400 mm diameter volume-of-interest. GD increased, as expected, with decreasing BW, and with the 2D versus 3D correction algorithm. Differences in GD attributed to phantom set-up were 0.13 mm or less. Differences in GD for the two software applications were less than 0.07 mm. A novel modular phantom was developed to evaluate distortions in MR images for radiation therapy applications. |
Databáze: |
MEDLINE |
Externí odkaz: |
|