Use of uniform shots for robust planning of mask-based treatment in Gamma Knife Icon.

Autor: Luo D; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA. Electronic address: dluo@mdanderson.org., Han EY; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA., Wen Z; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA., Mackin DS; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA., Krafft SP; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA., Li J; Department of Radiation Oncology, Unit 97, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA., Wang X; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA., Briere TM; Department of Radiation Physics, Unit 94, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA.
Jazyk: angličtina
Zdroj: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2020 May; Vol. 73, pp. 135-157. Date of Electronic Publication: 2020 Apr 29.
DOI: 10.1016/j.ejmp.2020.04.013
Abstrakt: Purpose: To verify whether Icon automatic correction is robust in preserving plan quality.
Materials/methods: An end-to-end phantom was used to verify Icon's correction accuracy qualitatively. For quantitative assessment, two plans, a composite- and a uniform-shot-only, were created for an elliptical- (E) and a sausage-shaped (S) lesion inside a PseudoPatient head phantom with a film insert. The phantom was irradiated in the planned and three other positions under each plan: 14° pitch (B); 14° rotation + 8° pitch (C); 95° rotation + 4-cm shift (D).
Results: Icon accurately corrects the locations of the shots. For the uniform-shot plans: all gamma index passing rates were >97%, and the differences between the planned and the delivery doses (minimum, maximum, and mean) were all ≤0.1 Gy. For the composite-shot plans, however, the dose differences increased as the phantom was shifted through positions B-D, with a gamma index passing rate of 61% for lesion-E in position D, and 92%, 79%, and 45% for lesion-S in positions B, C, and D, respectively.
Conclusions: Plans using only uniform shots are more robust to deviations in treatment position. The tolerance for such deviations may be lower for plans using composite shots.
(Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.)
Databáze: MEDLINE