Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display.
Autor: | Alawa KA; Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Nolan RP; Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Han E; Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Arboleda A; Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Durkee H; Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Sayed MS; Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Aguilar MC; Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA., Lee RK; Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA RLee@med.miami.edu. |
---|---|
Jazyk: | angličtina |
Zdroj: | The British journal of ophthalmology [Br J Ophthalmol] 2021 Mar; Vol. 105 (3), pp. 440-444. Date of Electronic Publication: 2019 Sep 17. |
DOI: | 10.1136/bjophthalmol-2019-314031 |
Abstrakt: | Background: Current visual field screening machines are bulky and expensive, limiting their accessibility, affordability and use. We report the design and evaluation of a novel, portable, cost-effective system for glaucoma screening using smartphone-based visual field screening using frequency doubling technology (FDT) and a head-mounted display. Methods: Nineteen eyes of 10 subjects with new-onset or chronic primary open angle glaucoma were tested and compared with the Humphrey Zeiss FDT and the newly developed Mobile Virtual Perimetry (MVP) FDT with the C-20 testing pattern. Mann-Whitney, Bland-Altman and linear regression analyses were performed to assess statistical difference, agreement and correlation, respectively, between the two devices. Results: The average age of the participants was 58±15 years. No statistically significant difference was found between the MVP FDT and the Humphrey Zeiss FDT (p>0.05). Bland-Altman and linear regression analyses demonstrated good agreement and correlation between the two devices. Conclusion: The MVP FDT is a low-cost, portable visual field screening device that produces comparable results to the Humphrey Zeiss FDT and may be used as an easily accessible screening tool for glaucoma. Competing Interests: Competing interests: None declared. (© Author(s) (or their employer(s)) 2021. No commercial re-use. See rights and permissions. Published by BMJ.) |
Databáze: | MEDLINE |
Externí odkaz: |