Evaluation of the force generated by gradual deflection of 0.016-inch NiTi and stainless steel orthodontic wires in self-ligating metallic and esthetic brackets.

Autor: Francisconi MF; D.D.S., M.Sc. Orthodontic Graduate Student. Department of Orthodontics. Bauru Dental School. University of São Paulo, Bauru, Brazil., Janson G; D.D.S., M.Sc., Ph.D. Professor and Head. Department of Orthodontics. Bauru Dental School, University of São Paulo, Bauru, Brazil., Henriques JF; D.D.S., M.Sc., Ph.D. Professor. Department of Orthodontics. Bauru Dental School, University of São Paulo, Bauru, Brazil., Freitas KM; D.D.S., M.Sc. PhD. Professor. Department of Orthodontics. UNINGA University Center, Maringa, Brazil., Francisconi PA; D.D.S., M.Sc., Ph.D. Associate Professor. Department of Operative Dentistry, Endodontics and Dental Materials. Bauru Dental School, University of São Paulo, Bauru, Brazil.
Jazyk: angličtina
Zdroj: Journal of clinical and experimental dentistry [J Clin Exp Dent] 2019 May 01; Vol. 11 (5), pp. e464-e469. Date of Electronic Publication: 2019 May 01 (Print Publication: 2019).
DOI: 10.4317/jced.55698
Abstrakt: Background: The purpose of this study was to evaluate the deflection forces of 0.016-inch Nitinol and stainless steel orthodontic wires, in association to different self-ligating brackets.
Material and Methods: Specimens were mounted in a clinical simulation model and evaluated in a Universal Testing Machine (INSTRON 3342), using a 10N load cell and ISO 15,841, as a protocol. Eight of these models were prepared, each one for the bonding of each set of self-ligating accessories to be tested: Damon Q, Damon Clear (Ormco), In-Ovation R, In-Ovation C (GAC), BioQuick, QuickClear (Forestadent), SmartClip and Clarity SL (3M). Data were subjected to One-way ANOVA, followed by Tukey tests ( P <0.05).
Results: Elastic deflection results showed that the deactivation forces increased with increase in wire deflection in the different brackets evaluated. For the different combinations, Clarity SL generated the greatest force and Damon Clear presented the lowest force when compared to the other brackets in all alloys and deflections. BioQuick and QuicKlear were those with the most similar behavior with each other.
Conclusions: Strength values increased with gradual increase in wire deflection in all evaluated brackets. Clarity SL generated the greatest and Damon Clear the lowest force when compared to the other brackets in all alloys and deflections tested. Key words: Brackets, orthodontic wires, deflection.