Relation between insertion torque and tactile, visual, and rescaled gray value measures of bone quality: a cross-sectional clinical study with short implants.
Autor: | Triches DF; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil. diego.triches@pucrs.br., Alonso FR; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil., Mezzomo LA; Postgraduate Program in Dentistry, Federal University of Santa Catarina (UFSC), Florianópolis, Brazil., Schneider DR; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil., Villarinho EA; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil., Rockenbach MI; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil., Teixeira ER; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil., Shinkai RS; Postgraduate Program in Dentistry, Pontifical Catholic University of Rio Grande do Sul (PUCRS), Dental School, Avenida Ipiranga, 6681 - Prédio 6, Porto Alegre, RS, 90619-900, Brazil. |
---|---|
Jazyk: | angličtina |
Zdroj: | International journal of implant dentistry [Int J Implant Dent] 2019 Feb 11; Vol. 5 (1), pp. 9. Date of Electronic Publication: 2019 Feb 11. |
DOI: | 10.1186/s40729-019-0158-6 |
Abstrakt: | Background: This study assessed the relationship between insertion torque and bone quality evaluated during surgery and in preoperative computed tomographic (CT) images analyzed either visually or by rescaled mean gray values (MGVs). The study also tested the correlation between the clinical and radiographic measures of bone quality. Methods: The consecutive sample was composed of 45 short implants (4.1 × 6 mm) placed in the posterior region of 20 patients. Intra-surgical tactile bone quality, based on the classification of bone types by Lekholm and Zarb, and insertion torque were recorded during the implant placement. Visual bone quality and normalized MGV were assessed in standardized axial, coronal, and sagittal sections of preoperative CT images. Data were analyzed by ANOVA and Spearman correlation (alpha = 0.05). Results: Insertion torque was associated with all assessment methods of bone quality (tactile, CT visual, MGV). A moderate correlation was found among all methods of bone quality, except for CT visual assessment and tactile evaluation. MGVs varied as a function of arch, dental region, insertion torque, and bone types. Conclusions: The results suggest that bone quality measures affect primary stability as recorded by insertion torque, and the assessment methods are consistently related. |
Databáze: | MEDLINE |
Externí odkaz: |