A Practical Approach to Electrode-Skin Impedance Unbalance Measurement.

Autor: Spinelli E, Mayosky MA, Pallàs-Areny R
Jazyk: angličtina
Zdroj: IEEE transactions on bio-medical engineering [IEEE Trans Biomed Eng] 2019 Jan 14. Date of Electronic Publication: 2019 Jan 14.
DOI: 10.1109/TBME.2006.873699
Abstrakt: Unbalance between electrode-skin impedances is a major problem in biopotential recordings, leading to increased power-line interference. This paper proposes a simple, direct method to measure that unbalance at power-line frequency (50-60Hz), thus allowing the determination of actual recording conditions for biopotential amplifiers. the method is useful in research, amplifier testing, electrode design and teaching purposes. It has been experimentally validated by using both phantom impedances and real electrode-skin impedances.
Databáze: MEDLINE