A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars.
Autor: | Piasecki L; Department of Periodontics and Endodontics, University at Buffalo, Buffalo, New York. Electronic address: lucilapi@buffalo.edu., José Dos Reis P; Department of Physics, State University of Central West, Guarapuava, Paraná, Brazil., Jussiani EI; Applied Nuclear Physics Laboratory, Physics Department, State University of Londrina, Londrina, Paraná, Brazil., Andrello AC; Applied Nuclear Physics Laboratory, Physics Department, State University of Londrina, Londrina, Paraná, Brazil. |
---|---|
Jazyk: | angličtina |
Zdroj: | Journal of endodontics [J Endod] 2018 Dec; Vol. 44 (12), pp. 1872-1877. Date of Electronic Publication: 2018 Nov 01. |
DOI: | 10.1016/j.joen.2018.09.001 |
Abstrakt: | Introduction: This study evaluated the accuracy of 3 electronic apex locators (EALs) (CanalPro [Coltene-Endo, Cuyahoga Falls, OH], Apex ID [SybronEndo, Glendora, CA], and Root ZX Mini [J Morita, Tokyo, Japan]) in curved mesial canals of extracted mandibular molars using micro-computed tomographic (micro-CT) scanning. Methods: The root canal length and the actual working length of 58 canals were measured using the visual method and 3-dimensional micro-CT reconstructions. The measurements of the EALs at marks "APEX/0.0" and "0.5 mark" were recorded as the electronic root canal length and the electronic working length, respectively. The absolute mean values and the percentages of distribution of the electronic measurements were compared with the actual lengths of the canals. The presence of isthmuses, accessory canals, foramen position, and the distance between the apical constriction and the apical foramen were recorded from the micro-CT scans, and their correlation to the accuracy of the EALs was determined. Results: All electronic measurements showed high agreement with their respective gold standard, except the electronic root canal length of the Apex ID (P < .05). No difference in the percentage of precise measurements (within the ±0.5 mm) was found among the devices (χ 2 and z tests, P > .05). Of the anatomic parameters evaluated, the presence of a lateral foramen negatively affected the 0.0 mark of the Apex ID (Wilcoxon test, P < .05). Conclusions: The Root ZX Mini and CanalPro were precise for both root canal length and working length determination in mesial curved canals of mandibular molars, whereas the Apex ID was accurate for the working length when using the 0.5 mark. (Copyright © 2018 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |