Effect of varying skin surface electrode position on electroretinogram responses recorded using a handheld stimulating and recording system.

Autor: Hobby AE; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.; Centre of Applied Vision Research, City, University of London, Northampton Square, London, EC1V 0HB, UK.; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK., Kozareva D; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK., Yonova-Doing E; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK., Hossain IT; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK., Katta M; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK., Huntjens B; Centre of Applied Vision Research, City, University of London, Northampton Square, London, EC1V 0HB, UK., Hammond CJ; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK.; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK., Binns AM; Centre of Applied Vision Research, City, University of London, Northampton Square, London, EC1V 0HB, UK., Mahroo OA; Department of Ophthalmology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK. omar.mahroo@moorfields.nhs.uk.; Department of Twin Research and Genetic Epidemiology, King's College London, St Thomas' Hospital Campus, London, SE1 7EH, UK. omar.mahroo@moorfields.nhs.uk.; UCL Institute of Ophthalmology, University College London, London, EC1V 9EL, UK. omar.mahroo@moorfields.nhs.uk.; Medical Retina Service, Moorfields Eye Hospital, 162 City Road, London, EC1V 2PD, UK. omar.mahroo@moorfields.nhs.uk.; Department of Physiology, Development and Neuroscience, University of Cambridge, Cambridge, CB2 3EG, UK. omar.mahroo@moorfields.nhs.uk.
Jazyk: angličtina
Zdroj: Documenta ophthalmologica. Advances in ophthalmology [Doc Ophthalmol] 2018 Oct; Vol. 137 (2), pp. 79-86. Date of Electronic Publication: 2018 Jul 25.
DOI: 10.1007/s10633-018-9652-z
Abstrakt: Purpose: A handheld device (the RETeval system, LKC Technologies) aims to increase the ease of electroretinogram (ERG) recording by using specially designed skin electrodes, rather than corneal electrodes. We explored effects of electrode position on response parameters recorded using this device.
Methods: Healthy adult twins were recruited from the TwinsUK cohort and underwent recording of light-adapted flicker ERGs (corresponding to international standard stimuli). In Group 1, skin electrodes were placed in a "comfortable" position, which was up to 20 mm below the lid margin. For subsequent participants (Group 2), the electrode was positioned 2 mm from the lid margin as recommended by the manufacturer. Amplitudes and peak times (averaged from both eyes) were compared between groups after age-matching and inclusion of only one twin per pair. Light-adapted flicker and flash ERGs were recorded for an additional 10 healthy subjects in two consecutive recording sessions: in the test eye, electrode position was varied from 2 to 10-20 mm below the lid margin between sessions; in the fellow (control) eye, the electrode was 2 mm below the lid margin throughout. Amplitudes and peak times (test eye normalised to control eye) were compared for the two sessions.
Results: Including one twin per pair, and age-matching yielded 28 individuals per group. Flicker ERG amplitudes were significantly lower for Group 1 than Group 2 participants (p = 0.0024). However, mean peak times did not differ between groups (p = 0.54). For the subjects in whom electrode position was changed between recording sessions, flash and flicker amplitudes were significantly lower when positioned further from the lid margin (p < 0.005), but peak times were similar (p > 0.5).
Conclusions: Moving the skin electrodes further from the lid margin significantly reduces response amplitudes, highlighting the importance of consistent electrode positioning. However, this does not significantly affect peak times. Thus, it may be feasible to adopt a more comfortable position in participants who cannot tolerate the recommended position if analysis is restricted to peak time parameters.
Databáze: MEDLINE