High-Volume-Rate 3-D Ultrasound Imaging Based on Synthetic Aperture Sequential Beamforming With Chirp-Coded Excitation.

Autor: Zhou J, Wei S, Jintamethasawat R, Sampson R, Kripfgans OD, Fowlkes JB, Wenisch TF, Chakrabarti C
Jazyk: angličtina
Zdroj: IEEE transactions on ultrasonics, ferroelectrics, and frequency control [IEEE Trans Ultrason Ferroelectr Freq Control] 2018 Aug; Vol. 65 (8), pp. 1346-1358. Date of Electronic Publication: 2018 May 21.
DOI: 10.1109/TUFFC.2018.2839085
Abstrakt: Three-dimensional (3-D) ultrasound imaging is a promising modality for many medical applications. Unfortunately, it generates voluminous data in the front end, making it unattractive for high-volume-rate portable medical applications. We apply synthetic aperture sequential beamforming (SASB) to greatly compress the front-end receive data. Baseline 3-D SASB has a low volume rate, because subapertures fire one by one. In this paper, we propose to increase the volume rate of 3-D SASB without degrading imaging quality through: 1) transmitting and receiving simultaneously with four subapertures and 2) using linear chirps as the excitation waveform to reduce interference. We design four linear chirps that operate on two overlapped frequency bands with chirp pairs in each band having opposite chirp rates. Direct implementation of this firing scheme results in grating lobes. Therefore, we design a sparse array that mitigates the grating lobe levels through optimizing the locations of transducer elements in the bin-based random array. Compared with the baseline 3-D SASB, the proposed method increases the volume rate from 8.56 to 34.2 volumes/s without increasing the front-end computation requirement. Field-II-based cyst simulations show that the proposed method achieves imaging quality comparable with baseline 3-D SASB in both shallow and deep regions.
Databáze: MEDLINE