Reliability and Validity of a 1-Person Technique to Measure Humeral Torsion Using Ultrasound.
Autor: | Hannah DC; Department of Orthopaedic Surgery, Augusta University, GA.; Rehabilitation Science Program., Scibek JS; Department of Athletic Training., Carcia CR; Department of Physical Therapy., Phelps AL; Palumbo Donahue School of Business, Duquesne University, Pittsburgh, PA. |
---|---|
Jazyk: | angličtina |
Zdroj: | Journal of athletic training [J Athl Train] 2018 Jun; Vol. 53 (6), pp. 590-596. Date of Electronic Publication: 2018 Jul 05. |
DOI: | 10.4085/1062-6050-213-17 |
Abstrakt: | Context: Knowledge of the bilateral difference in humeral torsion (HT) enables clinicians to implement appropriate interventions for soft tissue restrictions of the shoulder to restore rotational motion and reduce injury risk. Whereas the current ultrasound method for measuring HT requires 2 assessors, a more efficient 1-person technique (1PT) may be of value. Objective: To determine if a 1PT is a reliable and valid alternative to the established 2-person technique (2PT) for indirectly measuring HT using ultrasound. Design: Descriptive laboratory study. Setting: Biomechanics laboratory. Patients or Other Participants: A convenience sample of 16 volunteers (7 men, 9 women; age = 26.9 ± 6.8 years, height = 172.2 ± 10.7 cm, mass = 80.0 ± 13.3 kg). Main Outcome Measure(s): We collected the HT data using both the 1PT and 2PT from a total of 30 upper extremities (16 left, 14 right). Within-session intrarater reliability (intraclass correlation coefficient; ICC [3,1]) and standard error of measurement (SEM) were assessed for both techniques. Simple linear regression and Bland-Altman analysis were used to examine the validity of the 1PT when compared with the established 2PT. Results: The 1PT (ICC [3,1] = 0.992, SEM = 0.8°) and 2PT (ICC [3,1] = 0.979, SEM = 1.1°) demonstrated excellent within-session intrarater reliability. A strong linear relationship was demonstrated between the HT measurements collected with both techniques ( r = 0.963, r 2 = 0.928, F Conclusions: The 1PT for measuring HT using ultrasound was a reliable and valid alternative to the 2PT. By reducing the number of testers involved, the 1PT may provide clinicians with a more efficient and practical means of obtaining these valuable clinical data. a . |
Databáze: | MEDLINE |
Externí odkaz: |