Dosimetric characterisation of the optically-stimulated luminescence dosimeter in cobalt-60 high dose rate brachytherapy system.

Autor: Rejab M; Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia., Wong JHD; Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia.; University of Malaya Research Imaging Centre, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia., Jamalludin Z; Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia., Jong WL; Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia., Malik RA; Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia., Wan Ishak WZ; Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia., Ung NM; Clinical Oncology Unit, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia. nm_ung@um.edu.my.
Jazyk: angličtina
Zdroj: Australasian physical & engineering sciences in medicine [Australas Phys Eng Sci Med] 2018 Jun; Vol. 41 (2), pp. 475-485. Date of Electronic Publication: 2018 May 13.
DOI: 10.1007/s13246-018-0647-6
Abstrakt: This study investigates the characteristics and application of the optically-stimulated luminescence dosimeter (OSLD) in cobalt-60 high dose rate (HDR) brachytherapy, and compares the results with the dosage produced by the treatment planning system (TPS). The OSLD characteristics comprised linearity, reproducibility, angular dependence, depth dependence, signal depletion, bleaching rate and cumulative dose measurement. A phantom verification exercise was also conducted using the Farmer ionisation chamber and in vivo diodes. The OSLD signal indicated a supralinear response (R 2  = 0.9998). It exhibited a depth-independent trend after a steep dose gradient region. The signal depletion per readout was negligible (0.02%), with expected deviation for angular dependence due to off-axis sensitive volume, ranging from 1 to 16%. The residual signal of the OSLDs after 1 day bleached was within 1.5%. The accumulated and bleached OSLD signals had a standard deviation of ± 0.78 and ± 0.18 Gy, respectively. The TPS was found to underestimate the measured doses with deviations of 5% in OSLD, 17% in the Farmer ionisation chamber, and 7 and 8% for bladder and rectal diode probes. Discrepancies can be due to the positional uncertainty in the high-dose gradient. This demonstrates a slight displacement of the organ at risk near the steep dose gradient region will result in a large dose uncertainty. This justifies the importance of in vivo measurements in cobalt-60 HDR brachytherapy.
Databáze: MEDLINE