Centrifugation-free washing: A novel approach for removing immunoglobulin A from stored red blood cells.

Autor: Vörös E; Department of Biomedical Engineering, University of Houston, Houston, Texas 77204., Piety NZ; Department of Biomedical Engineering, University of Houston, Houston, Texas 77204., Strachan BC; Department of Biomedical Engineering, University of Houston, Houston, Texas 77204., Lu M; Department of Biomedical Engineering, University of Houston, Houston, Texas 77204., Shevkoplyas SS; Department of Biomedical Engineering, University of Houston, Houston, Texas 77204.
Jazyk: angličtina
Zdroj: American journal of hematology [Am J Hematol] 2018 Aug; Vol. 93 (4), pp. 518-526. Date of Electronic Publication: 2018 Jan 25.
DOI: 10.1002/ajh.25026
Abstrakt: Washed red blood cells (RBCs) are indicated for immunoglobulin A (IgA) deficient recipients. Centrifugation-based cell processors commonly used by hospital blood banks cannot consistently reduce IgA below the recommended levels, hence double washing is frequently required. Here, we describe a prototype of a simple, portable, disposable system capable of washing stored RBCs without centrifugation, while reducing IgA below 0.05 mg/dL in a single run. Samples from RBC units (n = 8, leukoreduced, 4-6 weeks storage duration) were diluted with normal saline to a hematocrit of 10%, and then washed using either the prototype washing system, or via conventional centrifugation. The efficiency of the two washing methods was quantified and compared by measuring several key in vitro quality metrics. The prototype of the washing system was able to process stored RBCs at a rate of 300 mL/hour, producing a suspension of washed RBCs with 43 ± 3% hematocrit and 86 ± 7% cell recovery. Overall, the two washing methods performed similarly for most measured parameters, lowering the concentration of free hemoglobin by >4-fold and total free protein by >10-fold. Importantly, the new washing system reduced the IgA level to 0.02 ± 0.01 mg/mL, a concentration 5-fold lower than that produced by conventional centrifugation. This proof-of-concept study showed that centrifugation may be unnecessary for washing stored RBCs. A simple, disposable, centrifugation-free washing system could be particularly useful in smaller medical facilities and resource limited settings that may lack access to centrifugation-based cell processors.
(© 2017 Wiley Periodicals, Inc.)
Databáze: MEDLINE