Autor: |
Rodríguez-Nuñez O, López-Téllez JM, Rodríguez-Herrera OG, Bruce NC |
Jazyk: |
angličtina |
Zdroj: |
Applied optics [Appl Opt] 2017 May 20; Vol. 56 (15), pp. 4398-4405. |
DOI: |
10.1364/AO.56.004398 |
Abstrakt: |
We present a method for calibration and data extraction for a nonoptimized Mueller matrix polarimeter. The advantage of this type of method is a reduction in measurement time for multiwavelength systems or in systems with slow response times. The calibration process requires the measurement of four known polarization devices. Here we use free-space transmission, a horizontal and a vertical linear polarizer, and a quarter-wave retarder with its fast axis at 30° to the horizontal. Experimental measurements of rotating quarter-wave and half-wave retarders show that accurate results can be obtained with the proposed method. |
Databáze: |
MEDLINE |
Externí odkaz: |
|