A BMI-adjusted ultra-low-dose CT angiography protocol for the peripheral arteries-Image quality, diagnostic accuracy and radiation exposure.
Autor: | Schreiner MM; Section of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. Electronic address: markus.schreiner@meduniwien.ac.at., Platzgummer H; Section of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. Electronic address: hannes.platzgummer@meduniwien.ac.at., Unterhumer S; Section of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. Electronic address: sylvia.unterhumer@meduniwien.ac.at., Weber M; Section of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. Electronic address: michael.weber@meduniwien.ac.at., Mistelbauer G; Institute of Computer Graphics and Algorithms, Technical University of Vienna, Favoritenstraße 9-11, 1040 Vienna, Austria. Electronic address: gabriel.mistelbauer@cg.tuwien.ac.at., Loewe C; Section of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. Electronic address: christian.loewe@meduniwien.ac.at., Schernthaner RE; Section of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. Electronic address: ruediger@schernthaner.eu. |
---|---|
Jazyk: | angličtina |
Zdroj: | European journal of radiology [Eur J Radiol] 2017 Aug; Vol. 93, pp. 149-156. Date of Electronic Publication: 2017 Jun 03. |
DOI: | 10.1016/j.ejrad.2017.06.002 |
Abstrakt: | Objectives: To investigate radiation exposure, objective image quality, and the diagnostic accuracy of a BMI-adjusted ultra-low-dose CT angiography (CTA) protocol for the assessment of peripheral arterial disease (PAD), with digital subtraction angiography (DSA) as the standard of reference. Methods: In this prospective, IRB-approved study, 40 PAD patients (30 male, mean age 72 years) underwent CTA on a dual-source CT scanner at 80kV tube voltage. The reference amplitude for tube current modulation was personalized based on the body mass index (BMI) with 120 mAs for [BMI≤25] or 150 mAs for [25 Results: The BMI-adjusted ultra-low-dose protocol reached high diagnostic accuracy values of 94% for Reader 1 and 93% for Reader 2. Moreover, in comparison to the ISP, it showed significantly (p<0.001) lower CTDIvol (1.97±0.55mGy vs. 4.18±0.62 mGy) and DLP (256±81mGy x cm vs. 544±83mGy x cm) but similar image quality (p=0.37 for CNR). Furthermore, image quality was similar between BMI groups (p=0.86 for CNR). Conclusions: A CT protocol that incorporates low kV settings with a personalized (BMI-adjusted) reference amplitude for tube current modulation and iterative reconstruction enables very low radiation exposure CTA, while maintaining good image quality and high diagnostic accuracy in the assessment of PAD. (Copyright © 2017 Elsevier B.V. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |