Antitachycardia pacing programming in implantable cardioverter defibrillator: A systematic review.

Autor: De Maria E; Elia De Maria, Ambra Borghi, Letizia Modonesi, Stefano Cappelli, Cardiology Unit, Ramazzini Hospital, 41012 Carpi (Modena), Italy., Giacopelli D; Elia De Maria, Ambra Borghi, Letizia Modonesi, Stefano Cappelli, Cardiology Unit, Ramazzini Hospital, 41012 Carpi (Modena), Italy., Borghi A; Elia De Maria, Ambra Borghi, Letizia Modonesi, Stefano Cappelli, Cardiology Unit, Ramazzini Hospital, 41012 Carpi (Modena), Italy., Modonesi L; Elia De Maria, Ambra Borghi, Letizia Modonesi, Stefano Cappelli, Cardiology Unit, Ramazzini Hospital, 41012 Carpi (Modena), Italy., Cappelli S; Elia De Maria, Ambra Borghi, Letizia Modonesi, Stefano Cappelli, Cardiology Unit, Ramazzini Hospital, 41012 Carpi (Modena), Italy.
Jazyk: angličtina
Zdroj: World journal of cardiology [World J Cardiol] 2017 May 26; Vol. 9 (5), pp. 429-436.
DOI: 10.4330/wjc.v9.i5.429
Abstrakt: Implantable cardioverter defibrillator (ICD) programming involves several parameters. In recent years antitachycardia pacing (ATP) has gained an increasing importance in the treatment of ventricular arrhythmias, whether slow or fast. It reduces the number of unnecessary and inappropriate shocks and improves both patient's quality of life and device longevity. There is no clear indication regarding the type of ATP to be used, except for the treatment of fast ventricular tachycardias (188 bpm-250 bpm) where it has been shown a greater efficacy and safety of burst compared to ramp; 8 impulses in each sequence of ATP appears to be the best programming option in this setting. Beyond ATP use, excellent clinical results were obtained with programming standardization following these principles: extended detection time in ventricular fibrillation (VF) zone; supraventricular discrimination criteria up to 200 bpm; first shock in VF zone at the maximum energy in order to reduce the risk of multiple shocks. The MADIT-RIT trial and some observational registries have also recently demonstrated that programming with a widespread use of ATP, higher cut-off rates or delayed intervention reduces the number of inappropriate and unnecessary therapies and improves the survival of patients during mid-term follow-up.
Competing Interests: Conflict-of-interest statement: Daniele Giacopelli is a Biotronik employee. The other authors report no relationships that could be construed as a conflict of interest.
Databáze: MEDLINE