Pitfalls in electrogram interpretation: Subcutaneous cardioverter defibrillator malfunction in Brugada syndrome.
Autor: | Morani G; Department of Cardiology, Azienda Ospedaliera Universitaria Integrata Verona, Verona, VR, Italy., Bolzan B; Department of Cardiology, Azienda Ospedaliera Universitaria Integrata Verona, Verona, VR, Italy., Tomasi L; Department of Cardiology, Azienda Ospedaliera Universitaria Integrata Verona, Verona, VR, Italy., Tomei R; Department of Cardiology, Azienda Ospedaliera Universitaria Integrata Verona, Verona, VR, Italy., Vassanelli C; Department of Cardiology, Azienda Ospedaliera Universitaria Integrata Verona, Verona, VR, Italy. |
---|---|
Jazyk: | angličtina |
Zdroj: | Pacing and clinical electrophysiology : PACE [Pacing Clin Electrophysiol] 2017 Oct; Vol. 40 (10), pp. 1180-1183. Date of Electronic Publication: 2017 May 16. |
DOI: | 10.1111/pace.13091 |
Abstrakt: | A patient with Brugada syndrome implanted with subcutaneous implantable cardioverter defibrillator (S-ICD) had oversensing episodes treated with S-ICD shocks. Comparable artifacts were not evocable with S-ICD pocket manipulation. The fluoroscopy excluded S-ICD macroscopic damage. The device extraction revealed undamaged pulse generator and connector, but the lead was inappropriately tunneled under the sixth rib. Then the S-ICD malfunction was due to lead microscopic damage caused by the lead rubbing the rib surface. (© 2017 Wiley Periodicals, Inc.) |
Databáze: | MEDLINE |
Externí odkaz: | |