Evaluation of the optimal cuff volume and cuff pressure of the revised laryngeal tube "LTS-D" in surgical patients.
Autor: | Kriege M; Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany., Alflen C; Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany., Eisel J; Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany., Ott T; Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany., Piepho T; Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany., Noppens RR; Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany. ruediger.noppens@lhsc.on.ca.; Department of Anesthesia & Perioperative Medicine, Western University; LHSC- University Hospital, 339 Windermere Road, London, ON, N6A 5A5, Canada. ruediger.noppens@lhsc.on.ca. |
---|---|
Jazyk: | angličtina |
Zdroj: | BMC anesthesiology [BMC Anesthesiol] 2017 Feb 02; Vol. 17 (1), pp. 19. Date of Electronic Publication: 2017 Feb 02. |
DOI: | 10.1186/s12871-017-0308-4 |
Abstrakt: | Background: Recent case reports have indicated significant cuff overinflation when using the standard filling volume based on the manufacturer's recommendations in older models of laryngeal tubes. The aim of this study was to determine the minimum cuff pressure needed to perform standardized ventilation without leakage in the new, revised model of the laryngeal tube "LTS-D". Methods: After ethical approval, LTS-D was placed for ventilation in 60 anesthetized patients. The cuff was inflated to the recommended volume (#3: 60 ml, #4: 80 ml, and #5: 90 ml). After evaluation of the initial cuff pressure (CP), the CP was lowered in 10 cmH Results: After initial inflation, the CP ranged from 105 cmH Conclusion: We found significant overinflation of the revised LTS-D using the recommended volume for initial cuff inflation. A CP of 60 cmH Trial Registration: ClinicalTrials.gov NCT02300337 . Registered: 20 November 2014. |
Databáze: | MEDLINE |
Externí odkaz: |