Intrasession and Intersession Repeatability of Diffusion Tensor Imaging in Healthy Human Liver.

Autor: Wong OL; From the *Department of Medical Physics and Applied Radiation Science, McMaster University, Hamilton, Ontario; †Imaging Research Center, St. Joseph's Healthcare, Hamilton Ontario, Canada; ‡Medical Physics and Research Department, §Comprehansive Oncology Center, and ∥Intervention and Radiology Department, Hong Kong Sanatorium & Hospital, Hong Kong Special Administrative Region (SAR), China; ¶School of Biomedical Engineering, and #Department of Electrical and Computer Engineering, McMaster University, Hamilton, Ontario, Canada., Leung TWT, Lo GG, Yuan J, Li WW, Noseworthy MD
Jazyk: angličtina
Zdroj: Journal of computer assisted tomography [J Comput Assist Tomogr] 2017 Jul/Aug; Vol. 41 (4), pp. 578-585.
DOI: 10.1097/RCT.0000000000000572
Abstrakt: Objective: The aim of this study was to evaluate the effect of signal to noise ratio (SNR) and number of gradient directions (NGD) on intra- and intersession repeatability of liver diffusion tensor imaging (DTI) metrics.
Methods: At each of 3 liver DTI scan sessions, liver diffusion was assessed in 5 healthy volunteers using a 6-direction DTI scan performed 9 separate times (ie, number of signal averages [NSA]). In addition, 4 combinations of NSA and NGD were acquired (NSA/NGD = 1/30, 3/10, 3/12, and 5/6) to determine the combined effect to DTI metrics, which was based on intersubject variability and intrasession (Vintra) and intersession (Vinter) repeatability.
Results: Intersubject variability was less than 20%, whereas Vintra and Vinter repeatability were less than 5% and less than 10%, respectfully. Vinter was not affected by the NGD used. Decreases in Vinter(FA), Vinter(λ1), Vinter(RD), and Vinter(MD) were observed with increasing NSA, and hence SNR.
Conclusion: Increased SNR may improve intrasession and intersession repeatability of liver DTI metrics. Scan repeatability was not influenced by NGD.